Thank you for choosing Apera Instruments SX721 Portable pH/ORP Meter (will be called “the meter” in 

short in the following content). Before using the meter, please read this instruction manual carefully in 

order to help you properly use and maintain it.

 

 

The  meter  can  measure  the  parameters  of  pH,  ORP,  and  temperature  of  water  solutions  with  high 

accuracy. It is suitable for various applications in different industries e.g., water treatment, aquaculture, 

horticulture, pools and spas, beverage making, mining, power plants, environmental monitoring, etc., 

especially ideal for in-field use.

 

 

The meter has the following features:

 

1.1.

 

Built-in  microprocessor  chip  with  the  intelligent  functions  of  auto.  calibration,  auto.  temperature 

compensation, auto. salinity compensation, manual baromatric pressure compensation, data storage, 

function settings, auto. power off, and low voltage display etc.

 

1.2. Adopts digital filter technology to intelligently improve meter’s response speed and accuracy. The 

smiley face icon will appear when the reading is stable.

 

1.3. Pure water modes available for pH, significantly increasing the accuracy in pure water measurement.

 

1.4. Meter’s LCD screen has clear and bright backlit display.

 

1.5. Built-tough. IP57 waterproof and dust-proof.

 

2  

Technical Specifications

 

2.1 

pH

 

Measurement range

 

(-2.00 to 19.99) pH

 

Resolution

 

0.1/0.01 pH

 

Accuracy

 

Meter: ±0.01pH; Overall: ±0.02pH

 

Input current

 

≤2×10

-12 

A

 

Input impendance

 

≥1×1012 Ω

 

Stability

 

±0.01 pH/3h

 

Temp. compensation range

 

(0 to 100) °C (automatic)

 

Summary of Contents for SX721

Page 1: ...1 SX721 Portable pH ORP Meter User Manual APERA Instruments Europe GmbH aperainst de...

Page 2: ...ctrode 7 6 2 pH Calibration 8 6 3 Notes about pH Calibration 9 6 4 Sample Test 9 6 5 Recommended pH Electrodes for Different Applications 10 6 6 Parameter Setting 11 6 7 Cleaning pH Electrode 13 6 8 T...

Page 3: ...oprocessor chip with the intelligent functions of auto calibration auto temperature compensation auto salinity compensation manual baromatric pressure compensation data storage function settings auto...

Page 4: ...r Supply Two AA alkaline batteries 1 5V x2 Dimensions and weight Meter 65 120 31 mm 180g Complete Kit 255 x 210x 50 mm 1380g Quality and safety certification CE ISO 9001 2015 2 4 Working Condition Wor...

Page 5: ...press to make changes 3 2 3 Function key a In measurement mode short press 1 5s to switch between and Long press to enter parameter setting P1 P2 P3 b In other measurement mode short press this key t...

Page 6: ...ickly review the data b In data recalling mode RM and serial number in the lower right corner short press to return to measurement mode 3 3 3 Delete data In data recalling mode hold for 5 seconds CLr...

Page 7: ...ompensation This electrode is only suitable for general water solutions pH testing Please refer to Section 6 5 for ideal pH electrodes to use for other applications A 3M KCL storage bottle comes with...

Page 8: ...stays on screen short press again to finish the 1st point calibration 7 00 pH will start flickering and a flickering c2 will show up indicating the 1st point calibration has been finished and the met...

Page 9: ...librated at 3 points 3 The frequency that you need to calibrate your meter depends on the tested samples condition of electrodes and the requirement of the accuracy For high accuracy measurements 0 02...

Page 10: ...bSen 803 Liquid food milk cream yogurt jam sauce etc LabSen 821 LabSen 823 High Temperature solutions LabSen 841 LabSen 861 Low temperature liquid LabSen881 High salinity solutions LabSen 841 Complex...

Page 11: ...e to producer setting OFF On 6 6 1 pH buffer solution series selection P1 a Long press meter enters P1 mode see picture 4 2 b Press or to choose buffer solution series U S A series 1 68 4 00 7 00 10 0...

Page 12: ...o choose temperature unit C or F c Press to enter next parameter setting or press to confirm and return to measurement mode 6 6 5 Back light display time setting P5 a Short press in P4 to enter P5 see...

Page 13: ...lution AI1166 for at least minutes Then use a soft brush to remove the contaminants Afterwards soak the electrode in 3M KCL soaking solution for at least 2 hours Rinse it off then re calibrate the tes...

Page 14: ...s to remove it The existence of an air bubble in the glass bulb will significantly decrease the stableness of measurement 3 pH electrode is technically a chemical battery Every pH electrode will event...

Page 15: ...llutant submerge the electrode in 0 1mol L dilute hydrochloric acid for 30 minutes then wash it in distilled water then submerge it in the soaking solution for 6 hours For organic or lipid pollutant c...

Page 16: ...y does not cover any damages due to Transportation storage improper use failure to follow the product instructions or to perform any preventive maintenance modifications combination or use with any pr...

Reviews: