6.6.7  Restore to factory default setting (P7) 

(a) Short press             in P6 to enter P7, see picture (4-8).

 

(b) Press           or           to choose         , meaning that the parameters 

have  been  restored  to  the  producer  setting  mode,  and  return  to 

measurement mode after 2 seconds. Returning to factory default setting 

is to restore the meter to theoretical value (zero potential pH is 7.00, slope 

is 100%), and set all the parameters to default settings. When the meter’s calibration or measurement is 

performing abnormally, users can use this function, and then calibrate and test again. Please note that 

this function is irreversible once used.

 

 

6.7 

Cleaning pH Electrode

 

1) 

The measurement is only as accurate as the electrode is clean. Always thoroughly rinse off the 

electrode before and after each test with pure water.

 

2) 

For tough contaminants, soak the electrode in Apera electrode cleaning solution (AI1166) for at 

least minutes. Then use a soft brush to remove the contaminants. Afterwards, soak the electrode 

in 3M KCL soaking solution for at least 2 hours. Rinse it off, then re-calibrate the tester before 

using again.

 

3) 

For other special contaminants, refer to the table below:

 

Contamination

 

Cleaning Solution

 

Inorganic metal oxide

 

Dilute acid less than 1mol/L

 

Organic lipid, protein containing samples

 

Apera cleaning solution (AI1166)

 

Resin macromolecule

 

Dilute alcohol, acetone, ether

 

Paints

 

Dilute bleacher, peroxide

 

 

6.8 

The Self-diagnose Information

 

During usage, the following icons may show up on LCD. This is the meter’s self-diagnose information, 

which can help you understand what may go wrong with the meter or electrode.

 

6.8.1.  The  stable  icon 

-2.00  pH

  or 

19.99  pH

  —  this  icon  appears  when  the  pH  value  exceeds  the 

measurement range. This icon will also show up when the electrode is not well connected with the meter 

or when the electrode is not insertd into solution. This is a normal phenomenon.

 

6.8.2.                — Electrode offset out of range (<-60mV or >60mV)

 

Summary of Contents for SX721

Page 1: ...1 SX721 Portable pH ORP Meter User Manual APERA Instruments Europe GmbH aperainst de...

Page 2: ...ctrode 7 6 2 pH Calibration 8 6 3 Notes about pH Calibration 9 6 4 Sample Test 9 6 5 Recommended pH Electrodes for Different Applications 10 6 6 Parameter Setting 11 6 7 Cleaning pH Electrode 13 6 8 T...

Page 3: ...oprocessor chip with the intelligent functions of auto calibration auto temperature compensation auto salinity compensation manual baromatric pressure compensation data storage function settings auto...

Page 4: ...r Supply Two AA alkaline batteries 1 5V x2 Dimensions and weight Meter 65 120 31 mm 180g Complete Kit 255 x 210x 50 mm 1380g Quality and safety certification CE ISO 9001 2015 2 4 Working Condition Wor...

Page 5: ...press to make changes 3 2 3 Function key a In measurement mode short press 1 5s to switch between and Long press to enter parameter setting P1 P2 P3 b In other measurement mode short press this key t...

Page 6: ...ickly review the data b In data recalling mode RM and serial number in the lower right corner short press to return to measurement mode 3 3 3 Delete data In data recalling mode hold for 5 seconds CLr...

Page 7: ...ompensation This electrode is only suitable for general water solutions pH testing Please refer to Section 6 5 for ideal pH electrodes to use for other applications A 3M KCL storage bottle comes with...

Page 8: ...stays on screen short press again to finish the 1st point calibration 7 00 pH will start flickering and a flickering c2 will show up indicating the 1st point calibration has been finished and the met...

Page 9: ...librated at 3 points 3 The frequency that you need to calibrate your meter depends on the tested samples condition of electrodes and the requirement of the accuracy For high accuracy measurements 0 02...

Page 10: ...bSen 803 Liquid food milk cream yogurt jam sauce etc LabSen 821 LabSen 823 High Temperature solutions LabSen 841 LabSen 861 Low temperature liquid LabSen881 High salinity solutions LabSen 841 Complex...

Page 11: ...e to producer setting OFF On 6 6 1 pH buffer solution series selection P1 a Long press meter enters P1 mode see picture 4 2 b Press or to choose buffer solution series U S A series 1 68 4 00 7 00 10 0...

Page 12: ...o choose temperature unit C or F c Press to enter next parameter setting or press to confirm and return to measurement mode 6 6 5 Back light display time setting P5 a Short press in P4 to enter P5 see...

Page 13: ...lution AI1166 for at least minutes Then use a soft brush to remove the contaminants Afterwards soak the electrode in 3M KCL soaking solution for at least 2 hours Rinse it off then re calibrate the tes...

Page 14: ...s to remove it The existence of an air bubble in the glass bulb will significantly decrease the stableness of measurement 3 pH electrode is technically a chemical battery Every pH electrode will event...

Page 15: ...llutant submerge the electrode in 0 1mol L dilute hydrochloric acid for 30 minutes then wash it in distilled water then submerge it in the soaking solution for 6 hours For organic or lipid pollutant c...

Page 16: ...y does not cover any damages due to Transportation storage improper use failure to follow the product instructions or to perform any preventive maintenance modifications combination or use with any pr...

Reviews: