Apera Instruments SX620 Instruction Manual Download Page 1

 

SX620 pH Pen Tester Instruction Manual 

 

1. Brief Introduction 
 

Thank you for purchasing APERA INSTRUMENTS SX620 pH Pen Tester. This product is featured with 

reliable and accurate measuring performance with a single-button operation in a slim body. Ideal for 

measuring pH in regular water solutions, including hydroponics, aquaculture, water treatment, pools& spas, 

cooling towers, beverage making, body fluid (e.g. urine, saliva). The pH probe is replaceable and suitable 

for test tube measuring and other small-volume sample testing. 

 

2. Specifications 

1)

  Measuring Range: -1 to 15.00 pH; Temperature: 0 to 60°C     

2)  

Resolution: 0.01 pH; Temperature: 0.1°C 

3)  

Accuracy: ± 0.01 pH, ±1 digit; Temperature: ± 0.5°C 

4)  

Automatic calibration: 1-3 point calibration (pH 7.00, pH4.00, pH10.01) 

5)

  Power: 2 x CR2032 Cell Batteries 

6)

  Dimension and Weight: 148*29*14mm / 43g 

7)

  Waterproof rating: IP57 

 

3. Instructions 

 

1) 

Calibration: Rinse the probe in distilled water and dap off excess water with tissue paper. Dip the probe 

in the standard buffer solution pH7.00. Stir gently, hold still, and wait for the reading to become stable. 

Then press the CAL button for about two seconds until 

CAL

 appears on the screen. When the 

calibration value 

7.0

 flashes, the calibration will finish after several seconds, and the meter will return to 

measurement mode and the icon 

will appear on the screen. 

2) 

Rinse the probe in distilled water and dap off excess water with tissue paper. Dip the probe in the other 

standard buffer solution (pH4.00 or pH10.01) to complete calibration according to the steps in 1). Icon 

 

will appear when pH 4.00 is calibrated, and icon 

 will appear when pH 10.01 is calibrated. 

3) 

Measurement: Rinse the probe in distilled water and dry it. Dip the probe in the sample solution. Stir gently 

and wait until you receive a stable reading. 

4) 

 After measurement, rinse the probe with distilled water, dry it, and put the protection cap back on. It’s a 

good practice to pour a bit storage solution that comes with kit into the protection cap when storing the 

meter, since it will help keep the accuracy of the pH probe. 

5) 

Temperature Unit Switch: When meter is turned off, hold the CAL button until the meter reboots and then 

the temperature unit will be switched (˚C–˚F) 

 

 

 