background image

 

- 7 - 

Symbol 

Self-Diagnosis information 

How to fix 

 

 

 

 

The  meter  cannot  recognize  the 

conductivity standard solutions. 

1. 

Make sure the probe is fully submerged in 
the solution. 

2. 

Check if the standard solution is expired or 
contaminated. 

3. 

Check if the conductivity electrode (two 
black rods) is damaged. 

4. 

Check if the conductivity electrode is 
contaminated. If so, please use a soft brush 
with warm water to clean up. 

5. 

If none of the above, please contact us at 

[email protected]

 or +1 (614) 285-3080 

 

 

 

        Is  pressed  before  measurement  is 

fully stable (          comes up and stays) 

Wait for          to come up and stays on screen before 

pressing         

 

During calibration, readings being unstable 

for over 3 minutes 

1. 

Shake the probe to remove the air bubbles 
on the surface of the black rods 

2. 

Check if the conductivity electrode is 
contaminated. If so, please use a soft brush 
with warm water to clean up. 

3. 

Soak the probe in 12.88mS/cm solution for 
10 minutes, then rinse with distilled water. 

4. 

If none of the above, please contact us at 

[email protected]

 or +1 (614) 285-3080 

 

The  calibration  reminder  is  triggered. 

It’s 

time  to  perform  a  new  conductivity 

calibration 

Perform  conductivity  calibration  or  cancel  calibration 

reminder in ZenTest settings. 

 

Conduct ivit y M easurement  

Press            key to turn on the tester. Press          to switch to Conductivity measuring mode. Rinse the probe 

in distilled water and remove excess water. Insert the probe in sample solution, shake it for a few seconds, 

and allow it to stand still in the solution until a stable reading is reached. Get readings after            comes up 

and stays. Press          to switch from conductivity to TDS, salinity, and resistivity. 

Summary of Contents for AI3718

Page 1: ... 1 SKU AI3718 EC60 Z Smart Multi Parameter Tester Conductivity TDS Salinity Resistivity Temp Instruction Manual APERA INSTRUMENTS LLC www aperainst com V1 03 ...

Page 2: ... 10 Warranty 11 ATTENTION 1 You may find a few drops of water in the probe cap These water droplets are added to maintain the sensitivity of the conductivity sensor before the product leaves factory It does NOT mean the product is used 2 The batteries are already preinstalled Just pull off the paper slip before using the tester When you replace the batteries make sure to follow the correct directi...

Page 3: ...splay dial display 3 Graph Mode digital display graph display 4 Table Mode digital display real time measurement and history display Calibration Press buttons to operate Operate on smartphone following graphic guides Self Diagnosis Er1 Er6 icons Detailed problem analysis and solutions Parameter Setup Press buttons to set up except for P7 and P11 All parameters can be set up in Settings Alarm The s...

Page 4: ...measurement mode long press to turn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting short press to change parameter Uni directional 1 Long press to enter calibration mode 2 In calibration mode short press to confirm calibration 3 In measurement mode when automatic lock is turned off short press to manually lock or unlock re...

Page 5: ...re stained gently clean the electrode with a soft brush in warm water containing detergent or alcohol 3 Things needed in addition to what s in the box a Distilled or deionized water 8 16oz for rinsing the probe after each test b Tissue paper for drying the probe Conductivity Calibration How to calibrate 1 Press key to switch to conductivity measurement mode Cond Rinse the probe in distilled water ...

Page 6: ...ffers we recommend that users replace new conductivity solutions after 5 to 10 times of use to keep the standard solution s accuracy Do NOT pour the used calibration solutions back into the solution bottles in case of contamination 5 Temperature compensation factor The default setting of the temperature compensation factor is 2 0 User can adjust the factor based on test solution and experimental d...

Page 7: ...r bubbles on the surface of the black rods 2 Check if the conductivity electrode is contaminated If so please use a soft brush with warm water to clean up 3 Soak the probe in 12 88mS cm solution for 10 minutes then rinse with distilled water 4 If none of the above please contact us at info aperainst com or 1 614 285 3080 The calibration reminder is triggered It s time to perform a new conductivity...

Page 8: ...automatic backlight time for 1 to 8 minutes For example if 3 minutes is set the backlight will turn off automatically after 3 minutes when the Off is set the auto backlight function will be turned off and short press to manually turn the backlight on or off 4 Auto Power off P04 The auto power off time can be set to 10 to 20 minutes For example if 15 minutes is set the meter will automatically shut...

Page 9: ...ing is cancelled in the ZenTest App the Er6 icon will disappear 6 Conductivity Back to Factory Default P8 Select Yes to recover instrument calibration to theoretical value This function can be used when instrument does not work well in calibration or measurement Calibrate and measure again after setting the instrument back to factory default Technical Specifications Conductivity Range 0 to 199 9 μ...

Page 10: ...at is compatible with EC60 Z is EC60 DE Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6 Stable reading indicator Waterproof Rating IP67 floats on water Reading Lock HOLD Power DC3V AAA batteries 4 Bluetooth Signal Battery Life 200 Hours Low power reminder Backlight White Measurement Green Calibration Red Alarm Auto Power Off Automatically power off if no operation for 10 minutes Di...

Page 11: ... to accidental damage unauthorized repair normal wear and tear or external causes such as accidents abuse or other actions or events beyond our reasonable control To claim for warranty or seek additional help please go to support aperainst com and click New support ticket Please provide the following information in the ticket Your Name Phone Number Email Shipping Address Product Name Serial Number...

Page 12: ... 12 APERA INSTRUMENTS LLC Address 6656 Busch Blvd Columbus Ohio 43229 Tel 1 614 285 3080 Email info aperainst com Website www aperainst com ...

Reviews: