8 Performing a Measurement
54
D52IB001EN-G
6. Take of the sample holder fixation ring and
replace it with the sample holder heating envi-
ronment.
7. Mount the sample holder heating environment
with the two screws.
8. Mount the sample holder as described in chap-
ter 8.4.1 Mounting the Standard Sample
Holder.
Nickel and graphite are available as foil materials for
the standard sample holder heating environment.
The choice depends mainly on the ROI (region of in-
terest) of the diffractogram. Background measure-
ments for Nickel and Graphite can be found in
chapter 8.3.
To replace the foil of the sample holder heating en-
vironment and to demount/mount the antiscatter
shield for the sample holder heating environment
proceed as follows:
1. Unscrew the two screws of the foil clamps and
take of the clamps and the foil
2. Mount/demount the antiscatter shield.
3. Replace the foil and ensure that the notches of
the foil are placed over the holes for the screws.
4. Mount the foil clamps with the two screws.
8.4.3 Mounting the Capillary Sample Holder
For information on mounting the capillary sample
holder, refer to the instruction manual Capillary
Sample Holder for TTK 600.