8 Performing a Measurement
D52IB001EN-G
45
2. Take out the heater insulation from the acces-
sory box delivered with the TTK 600 and place it
over the heater. See Fig. 9for further details.
3. Mount the heating block by reversing the actions
from point 1.
Fig. 49: Sample on standard sample holder
8.2.2 Additional Information for the Trans-
mission Sample Holder
The transmission sample holder (1) of TTK 600 is
made of nickel providing a good thermal conductivi-
ty and mechanical properties as well as high corro-
sion resistance.
A Pt100 sheath resistance temperature sensor (2) is
encapsulated in the sample holder (1), providing ex-
act temperature measurement. A connector for the
Pt100 (3) permits fast and easy exchange of the
sample holder. An additional convection heating is
provided by the connection of the gas hose (4) to the
gas connector on the right side of the instrument (5).
This additional heating mechanism ensure a minimi-
zation of the temperature deviation between dis-
played and real sample temperature. To protect the
detector, use the beamstop (6) at low 2
Θ
angles.
Fig. 50: Sample on transmission sample holder
8.2.3 Samples for the Capillary Sample
Holder
For information on preparing samples for the capil-
lary sample holder, refer to the instruction manual
Capillary Sample Holder for TTK 600.
NOTICE
Risk of damage.
Take care that the graphite foil that provides
improved thermal contact between sample
holder and heating/cooling block is in a good
condition (no visible scratches). Otherwise it is
strongly recommended to replace the graphite
foil with a new one in order to minimize the tem-
perature deviation between displayed and real
sample temperature.
1
2
3
NOTICE
Risk of damage.
Take care that the graphite foil that provides
improved thermal contact between sample
holder and heating/cooling block is in a good
condition (no visible scratches). Otherwise it is
strongly recommended to replace the graphite
foil with a new one in order to minimize the tem-
perature deviation between displayed and real
sample temperature.
Detailed information on the replacement of the
graphite foil are given in the service manual of
TTK 600.
1
2
3
4
5
6
Summary of Contents for TTK 600
Page 2: ......
Page 74: ...Appendix B Sample Chamber Dimensions 74 D52IB001EN G Appendix B Sample Chamber Dimensions ...
Page 75: ...Appendix C Declaration of Conformity D52IB001EN G 75 Appendix C Declaration of Conformity ...
Page 77: ...Appendix D Spare Parts and Accessories D52IB001EN G 77 176383 COOLING MODE SWITCHING VALVE ...