6 Operating CCU 100
40
D52IB001EN-G
6.3.1 Navigation Diagram of CCU 100
The complete navigation diagram is shown below:
Fig. 47: Navigation diagram
The navigation diagram of CCU 100 contains three
different branches. The first, which contains the pro-
cess value, the set point and the error messages is
already described in the previous chapter.
By pressing the SELECT button it is possible to
switch between the branches of the navigation dia-
gram. Pressing the button once, gives access to the
parameter page.
Parameter Page
The following parameters can be found in the pa-
rameter page:
Changing of the values within the parameter page is
possible by using the UP or DOWN buttons.
Config Parameters
Pressing the SELECT button twice gives access to
the config mode of the instrument. Parameters that
can be changed in the config mode are described in
the following table.
Parameter
Description
Note
rAte
Heating/Cool-
ing Rate value
[°C/min]
Default value:
20 °C/min
cOn
cOFF
Allows to switch
the cooling of
the instrument
on and off
Ln2
CAir
noC
Toggle between
the CCU 100
configurations
for LN2 cooling,
compressed-air
cooling and no
cooling
CL
Temperature
below which
cooling is per-
manently acti-
vated. Values
can be set
between 20 °C
and 50 °C.
Default value:
24 °C
rot.S
Sample holder
rotation activa-
tion;
values between
0 (no rotation)
and 10 (full
rotation speed)
can be adjusted
Only for mea-
surements with
the capillary
sample holder
of TTK 600
OP
shows the
actual heating
power of the
system in %.
Parameter
Description
Note
rAte: On
Off
Allows to deac-
tivate the heat-
ing rate. This
results in max.
heating power
all the time.
This parame-
ter is pass-
word
protected.
drAt
Allows changing
of the default
heating/cooling
rate (20 °C/min)
from 1 - 300 °C
This parame-
ter is pass-
word
protected.
Parameter
Description
Note
Summary of Contents for TTK 600
Page 2: ......
Page 74: ...Appendix B Sample Chamber Dimensions 74 D52IB001EN G Appendix B Sample Chamber Dimensions ...
Page 75: ...Appendix C Declaration of Conformity D52IB001EN G 75 Appendix C Declaration of Conformity ...
Page 77: ...Appendix D Spare Parts and Accessories D52IB001EN G 77 176383 COOLING MODE SWITCHING VALVE ...