2 TTK 600 - An Overview
16
D52IB001EN-G
Transmission:
Fig. 14: Battery Sample Holder (Transmission)
1 Fixing screw for sample holder
2 Pt100 connector
3 Lid of battery sample holder
4 Sheath resistance Pt100
5 Electrode cable
2.4 Sample Cooling
TTK 600 Low-Temperature Chamber can be oper-
ated with two types of sample cooling equipment,
depending on the desired temperature range:
• liquid nitrogen cooling
temperature range -190 °C to +600 °C
• compressed-air cooling
temperature range -20 °C to +600 °C
Both types of cooling equipment are connected to
TTK 600 via the connector for sample cooling (see
chapter 2.3.2) and controlled by CCU 100.
CCU 100 automatically turns off the cooling when it
is not needed to save liquid nitrogen/compressed-
air.
2.4.1 Liquid Nitrogen Cooling Equipment
The LN2 cooling equipment consists of the following
components:
• Low Temperature Equipment for Venturi Nozzle
(cat. no. 164578)
• Liquid Nitrogen Container, Dewar 60 L
(cat.no.166462)
or
• Liquid Nitrogen Container, Dewar 100 L
(cat.no.166463)
Other dewar vessels can be used, provided they
have the same specification and opening dimension
as the ones offered by Anton Paar GmbH (see
chapter 4.2.4 for further information).
Fig. 15: Low Temperature Equipment mounted on
TTK 600
1 LN2 connector
2 LN2 supply hose
3 Exhaust hose for evaporated nitrogen
Fig. 16: LN2 Dewar
1 LN2 Dewar vessel
2 LN2 inlet connection
3
4
1
5
2
1
2
3
2
1
Summary of Contents for TTK 600
Page 2: ......
Page 74: ...Appendix B Sample Chamber Dimensions 74 D52IB001EN G Appendix B Sample Chamber Dimensions ...
Page 75: ...Appendix C Declaration of Conformity D52IB001EN G 75 Appendix C Declaration of Conformity ...
Page 77: ...Appendix D Spare Parts and Accessories D52IB001EN G 77 176383 COOLING MODE SWITCHING VALVE ...