Lovis 2000 M/ME Low Volume – SOP 

Standard Operating Procedure for Low Sample Volume

9

Cleaning

TIP:  Select a suitable cleaning liquid. It must completely dissolve the sample, evaporate easily and 

without residues. It must not react chemically with the capillary or any of the wetted parts. 

If necessary, repeat all steps with a second cleaning liquid. 

Consult the Lovis 2000 M/ME Instruction Manual for more information.

1.   Unscrew the Screw Cap Front Side Closed from the Lovis capillary block.

2.   Pull out the capillary by the Filling Adapter Luer Female. 

3.   Hold the capillary with the tip upwards and remove the hollow plug from the tip.

4.   Turn the capillary around and hold it over a waste vessel. 

TIP:

To reuse the ball, carefully empty the sample residues plus ball into a clean vessel first. Depending 
on the sample’s properties the ball may not roll out. Remove it later when rinsing the capillary.

To clean the ball see 

"Cleaning the ball", p. 9

.

5.   Fill a clean syringe with a hollow needle with cleaning liquid. Insert the hollow needle into the 

capillary as far as possible. Fill the capillary with cleaning liquid. It should not overflow.

6.   If necessary wait a bit for the cleaning liquid to mix with the sample.

7.   Move the plunger up and down rapidly to improve the cleaning effect. Draw the cleaning liquid 

back into the syringe. Empty the syringe into a waste vessel.

8.   Repeat the two above steps two to three times until no traces of sample are left in the capillary. If 

the ball is still inside, let it roll out and rinse the capillary without the ball once.

9.   If necessary, repeat the cleaning procedure with a second cleaning liquid. 

10.   Wipe the capillary on the outside with some cleaning liquid and tissue paper.

11.   Put the hollow plug into some cleaning liquid. 

12.   Fill a clean syringe (5) with a hollow needle (6) with cleaning liquid. Insert the 

hollow needle into the plug (7) and rinse the inside of the hollow plug.

Suitable diameters: Hollow plug 1.59 or 1.8 >> Hollow needle 0.7 (gauge 22)

 Hollow plug 2.5

>> Hollow needle 1.2 (gauge 18)

13.   Take the hollow plug and wipe it with tissue paper.

14.   Blow the inside dry using a hollow needle and the built-in air pump. 

See below (how to dry the short capillary). 

15.   Blow the capillary (2) dry with the built-in air pump: 

Plug a clean hollow needle 1.2 (gauge 22) onto the Adapter Luer cone (13) of the air hose.

2 ....  Capillary

6 ....  Hollow needle
13....  Adapter Luer cone of air hose

Insert the hollow needle into the 
capillary as far as possible. 

Start the air pump with the button 

 on the touch screen.

The drying time depends on the cleaning liquid, on the temperature and on the ambient air 
humidity. Determine the suitable drying time empirically. 

Stop the pump with the button 

.

5

6

7

6

13

2

6

Summary of Contents for Lovis 2000 M/ME

Page 1: ...lary dry with the built in air pump Plug a clean hollow needle 1 2 gauge 18 onto the Adapter Luer cone 13 of the air hose 2 Capillary 6 Hollow needle 13 Adapter Luer cone of air hose Insert the hollow needle into the capillary as far as possible Start the air pump with the button on the touch screen The drying time depends on the cleaning liquid on the temperature and on the ambient air humidity D...

Page 2: ...r Lovis beforehand Tap OK The available Drive is displayed Select a capillary ball combination or a list of combinations in the field Target Tap Next If there is a list select one of the combinations You can assign a new name to the combination Select Create New Capillary if no capillary ball combination data is available Tap OK Choose ball type capillary diameter with extension short and enter th...

Page 3: ...d 7 Slowly fill the sample by pushing the plunger downwards 8 While the liquid rises draw the hollow needle upwards Make sure it remains below the liquid level and below any air bubbles 9 Fill sample until the liquid reaches the tip of the capillary 10 Check the entire filling The capillary must be filled without any bubbles or particles If detecting a bubble see arrow wait till it rises to the to...

Page 4: ...hem with the hollow needle 12 Check the filling again The capillary must be filled without any bubbles or particles 13 Select a suitable clean hollow plug 7 for the capillary 2 and close it Press the hollow plug firmly into the capillary Make sure not to insert air bubbles with the plug Turn the hollow plug to insert it deeper into the capillary If necessary remove a little sample with the hollow ...

Page 5: ... filling of the capillary for bubbles and check that the Lovis stands in a stable position Refill the capillary and tap Improve Lovis suggests Apply if the Level Adjustment correction is 0 5 The Level Adjustment is valid and stored after tapping Apply 6 Adjusting Lovis Measuring Systems Manager and Administrator only TIP If you have ordered a factory adjustment import the adjustment from the USB f...

Page 6: ...ion between adjustment and verification default 0 5 7 Fill the capillary once again with your standard liquid Do NOT change the ball Follow the instructions on the display When the verification is finished the runtime deviations for each adjusted angle are displayed TIP After a verification the runtime deviations to the values of the preceding original adjustment are displayed Typically these valu...

Page 7: ...surement is successfully completed if the progress bar turns green and the message above reads Finished The Lovis Condition is valid Menu Item Lovis Module Parameters Default Value Temperature Settings Temperature Measuring Temperature 20 C Temperature Equilibration This is the time Lovis waits for the sample to reach the measuring temperature after starting measurement and after the capillary blo...

Page 8: ...mple 7 Move the plunger up and down rapidly to improve the cleaning effect Draw the cleaning liquid back into the syringe Empty the syringe into a waste vessel 8 Repeat the two above steps two to three times until no traces of sample are left in the capillary If the ball is still inside let it roll out and rinse the capillary without the ball once 9 If necessary repeat the cleaning procedure with ...

Page 9: ...r only as far out as the µL scale Do not use strongalkaline cleaning liquids that will corrode the glass 1 Empty the syringe into a waste vessel 2 Draw in some cleaning liquid through the hollow needle and move the plunger up and down repeatedly If necessary wait a bit for the cleaning liquid to mix with the sample Empty the syringe into a waste vessel 3 Repeat the above step two to three times un...

Reviews: