6 Operating CCU 1000
B40IB006EN-D
29
6.3.1 Navigation Diagram of CCU 1000
The complete navigation diagram is shown below:
Fig. 18: Navigation diagram
The navigation diagram of CCU 1000 contains three
different branches. The first, which contains the pro-
cess value, the set point and the error messages is
already described in the previous chapter.
By pressing the SELECT button it is possible to
switch between the branches of the navigation dia-
gram. Pressing the button once, gives access to the
parameter page.
Parameter Page
The following parameters can be found in the pa-
rameter page:
Changing of the values within the parameter page is
possible by using the UP or DOWN buttons.
Config Parameters
Pressing the SELECT button twice gives access to
the config mode of the instrument. Parameters that
can be changed in the config mode are described in
the following table.
Parameter
Description
Note
rAte
Heating/Cool-
ing Rate value
[°C/min]
Default value:
20 °C/min
rEF
cAP
Configuration
for reflection or
for use with
capillary exten-
sion. Limits the
operating tem-
perature
depending on
the used equip-
ment.
Reflection:
max. 1200 °C
Capillary
Extension:
max. 1000 °C
24
Par
Conf
25
3E04
rate
ref
cap
rot.s
rate
drat
dSP
PROCESS VALUE
SET POINT
ERROR
PROCESS VALUE
SET POINT
ERROR
PROCESS VALUE
SET POINT
ERROR
code protected
read only
read only
on
off
0 – 10
0 … off
10 … full speed
code protected
code protected
OP
read only
rot.S
Sample holder
rotation activa-
tion;
values between
0 (no rotation)
and 10 (full
rotation speed)
can be adjusted
Only for mea-
surements with
the capillary
sample holder
of HTK 1200N
OP
shows the
actual heating
power of the
system in %.
Parameter
Description
Note
rAte: On
Off
Allows to deac-
tivate the heat-
ing rate. This
results in max.
heating power
all the time.
This parame-
ter is pass-
word
protected.
drAt
Allows changing
of the default
heating/cooling
rate (20 °C/min)
from 1 - 150 °C
This parame-
ter is pass-
word
protected.
dSP
Allows changing
of the default
set point (25 °C)
from 25 - 50 °C.
This parame-
ter is pass-
word
protected.
Parameter
Description
Note