13 Measuring
72
C76IB003EN-E
The minimum step size in temperature scans depends on the defined type of
equilibrium (Measurement finished by) of the density module:
13.2 Using the "No Sample List" Mode
If no sample changer or a sampler changer without magazine is connected, the
simple mode is set by default (see chapter 10.3). You only need to tap <OK> to
start a measurement.
You can use the quick settings function to change the following:
•
Sample name and measurement type
•
Measurement settings of the following measurement(s) without affecting
the settings of the underlying method (see Chapter 12.2.7)
To change the sample settings
1. Tap <Quick Settings>.
2. Enter a sample name using the corresponding input field if needed.
3. Define your parameters if needed.
4. Tap <OK>.
13.3 Using the "Sample List" Mode
The sample list is a tool that helps you managing samples that you want to
measure.
•
No results are stored in the sample list but using the <Single Sample>
button in the sample list links you to the data memory and you can view the
results of already measured samples in the sample list.
•
You can define temperature scans in the sample list.
•
Checks can also be started from the
"Checks/Adjustments"
menu.
•
Standard measurements can also be started from the main screen.
Measurement finished by
Minimum step size
Model
Predetermined
1 °C
DMA 4100/4500/5000 M
Equilibrium
0.02 °C
DMA 4100/4500 M
Equilibrium fast
Equilibrium medium
Equilibrium slow
0.02 °C
0.01 °C
0.005 °C
DMA 5000 M
NOTICE
The new settings are maintained for following measurements until changes
are made anew.