17 Service Utilities
C76IB003EN-E
103
2. Use the drop-down box "Group", select the type and tap <Next>.
3. Enter the "Values" for the Input Quantities and tap <Next>.
4. Read out the calculated results.
5. To calculate further results with other values, tap <Previous> and repeat
steps 3 and 4.
6. To exit the calculator, tap <OK>.
17 Service Utilities
In this chapter, you can find information about making a backup/restore of your
instrument settings, performing a firmware upgrade, getting details about the
system and viewing the live raw data.
17.1 Making a Backup of the Instrument Settings
You can use this utility for:
•
Saving the current instrument status including all method settings and your
favorites list for safety reasons for the case that someone changes the
settings accidentally.
•
Saving the current settings before you perform a firmware upgrade.
•
Copying the instrument settings to one or more other M series DMA
instruments.
A backup file contains the following instrument settings and instrument
information:
•
Global settings in the control panel (regional settings, keyboard settings,
network settings)
•
Method settings (display layout and content, result output settings, limit
settings, module settings, check definitions)
•
User functions
•
Users
•
System settings (instrument settings, global module settings, sample list
settings, audit trail settings, printout settings)
•
Adjustments (special adjustments, adjustment data)
To make a backup of instrument settings
1. Tap <Menu> and select
"Service > Backup Instrument Settings"
to open
the "Backup" dialog.
2. Select the data storage location (USB flash drive) and enter a "file name" or
accept the automatic file name.
3. Tap <Next>.