B46IB01
33
EN
10 Readjustment
•
Proper cleaning of the measuring cell is required before and after each series of
measurements.
•
Poor cleaning can result in the formation of deposits in the measuring cell. This
can cause offsets of the density readings, and can be recognized by measuring
distilled water.
•
If the density value deviates more than ±0.001 g/cm
3
from the theoretical value
(see Appendix B), then a readjustment may be necessary.
•
A readjustment is only possible if the density difference between measured and
theoretical value is below 0.100 g/cm
3
, and the water temperature is between 15
and 25 °C.
How to readjust:
1. Clean the measuring cell with a suitable cleaning agent until no more residues or
deposits are visible.
2. Remove the cleaning agent completely from the measuring cell.
3. Fill distilled water of approx. 20 °C into the measuring cell, carefully avoiding the
introduction of bubbles. The ambient temperature should be between 15 °C and
25 °C.
4. Turn the instrument off. Turn the instrument on again, pushing both the "ON/
OFF" key and the "Recall/Delete" key at the same time.
5. The measured density and "ADJ" are displayed.
The water temperature has to be between 15 and 25 °C, the density must be
between 0.900 and 1.100 g/cm
3
.
Summary of Contents for DMA 35N
Page 1: ...B46IB01 3 EN DMA 35N Portable Density Meter Firmware Version V2 014 Instruction Manual ...
Page 39: ...B46IB01 41 EN Appendix C CE Declaration of Conformity ...
Page 40: ...42 B46IB01 EN Appendix D EC Type Examination Certificate ...
Page 41: ...B46IB01 43 EN ...
Page 42: ...44 B46IB01 EN ...
Page 43: ...B46IB01 45 DE DMA 35N Tragbares Dichtemessgerät Firmwareversion V2 014 Betriebsanleitung ...
Page 81: ...B46IB01 83 DE Anhang C CE Konformitätserklärung ...
Page 82: ...84 B46IB01 DE Anhang D EG Baumuster Prüfbescheinigung ...
Page 83: ...B46IB01 85 DE ...
Page 84: ...86 B46IB01 DE ...
Page 85: ...B46IB01 87 FR DMA 35N Densimètre portatif Version du logiciel V2 014 Mode d emploi ...
Page 123: ...B46IB01 125 FR Annexe C Déclaration de conformité CE ...
Page 124: ...126 B46IB01 FR Annexe D Certificat d examen de type CE ...
Page 125: ...B46IB01 127 FR ...
Page 126: ...128 B46IB01 FR ...
Page 127: ...B46IB01 129 ES DMA 35N Densímetro Portátil Versión de Firmware V2 014 Manual de Instrucciones ...
Page 165: ...B46IB01 167 ES Apéndice C Declaración de conformidad CE ...
Page 166: ...168 B46IB01 ES Apéndice D Certificado de examen tipo EC ...
Page 167: ...B46IB01 169 ES ...
Page 168: ...170 B46IB01 ES ...