9.5.
CATTANI surgical separator
At the beginning of each work day.
Insert inside filter (
d
) a tablet (
v
) of VF CONTROL PLUS (CEFLA s.c.)
WARNING!
Always wear gloves to prevent contact with infected material when
carrying out this operation.
At the end of each surgical procedure.
• Execute an automatic flushing cycle or aspirate about ½ litre of sanitizing
solution with each of the cannulas used.
• Sterilize the cannula holder terminals in steam autoclave (see paragraph
1.5.).
At the end of each work day.
• Draw in 1 liter of water with each suction tube, alternating water and air
(keep the suction tube alternately in and out of the water).
• After rinsing with water, execute an automatic flushing cycle or aspirate
about ½ litre of sanitizing solution with each of the cannulas used.
NOTE:
After these operations, it is advisable to aspirate only air in
order to dry the entire suction system (5 minutes).
Every 15 days.
• Clean the separator container and probes with a soft sponge and neutral
detergent.
• Clean the drain valve for the separator’s container with the device provided
for this purpose.
Once a year.
• By technician: check the siphons and drains, check all the internal tubes and
plastic and rubber parts subject to wear.
Before leaving the surgery empty for a few days.
• Start the aspirator and run it 20 - 30 minutes without sucking in liquids.
The aspirator will dry itself completely. As a result, salt caused by moisture
and basic substances will not form, salt that may cause fan seizure and motor
blockage.
How to remove the separator’s container.
WARNING!
Gloves must be worn when carrying out the following operation to prevent
contact with infected material.
• Move the dental chair fully up.
• Open the side hydrogroup cover (see paragraph 7.7.).
•
Turn the electrical box
.
• Completely empty the separator bowl, pressing the timed button (
c
) located
on the cover.
• If present, remove the valve (
s
) for central systems.
•
Turn and raise the container (
d
) until it is detached from the drain pump (
k
).
• Detach the bowl (
d
) from the cover (
f
) pulling up the two side rubber bands
(
e
).
• After the cleaning operations, refit the vessel (
d
) after lubricating the
O-rings with S1-Protective Lubricant for O-Rings (CEFLA s.c.)
O-ring
before hand with silicon spray.
• Lastly, close the side hydrogroup cover.
Drain pump locked warning.
A dedicated icon (
A
) on the TOUCH DISPLAY will indicate if the drain
pump below the separator vessel locks.
At this point, shut off the equipment and empty the separator bowl by hand.
If the icon appears again, call technical service.
Summary of Contents for SMART TOUCH R7
Page 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Page 2: ......
Page 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Page 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Page 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Page 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Page 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Page 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Page 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Page 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Page 90: ......