7.6.
ACVS system (Automatic Cleaning Vacuum System)
Description of the system.
This system allows cleaning the surgical suction system.
The system comes with a tank (
c
) that contains the liquid disinfectant and
two fittings (
d
) used to wash the suction tubes.
The detergent liquid tank has a total capacity of 250 cc.
The washing cycle is automatically carried out and should usually be
performed at the end of each surgical procedure and whenever the dental
unit is cleaned and disinfected.
WARNING!
It is recommended to use STER 3 PLUS (CEFLA s.c.) as detergent liquid,
diluted in a 6% solution (equivalent to 60 ml of product in 1 litre of water).
How to start the washing cycle.
To start the washing cycle, follow the directions given below:
• Check that the tank (
c
) contains the detergent liquid.
• Remove both suction tube terminals from the assistant’s board, making
sure the suction motor starts running.
• Open the mechanical closing of the suction tube terminals.
• Insert the terminals in the fittings (
d
) found under the manifold.
The vacuum created by the Venutri meters triggers the washing cycle.
• Washing cycle stages:
- deliver municipal water for 50 sec. using intermittent operation (2 sec.
ON - 1 sec. OFF):
- stop the water flow and let in 10 cc of liquid disinfectant;
- stop letting in liquid disinfectant and continue sucking for 10 sec.
• The washing cycle ends when the suction flow is interrupted and the
motor stops running.
“Put the suction tubes back in place” appears on the display.
• At this point, put the ends of the suction tubes in the supports on the
assistant’s board to go back to the work conditions.
Filling the tank.
If the detergent liquid in the tank (
c
) is below the minimum level, act as
follows:
• Position the dental chair at maximum height.
• Remove the tank by turning it anticlockwise.
• Pour the detergent liquid into the tank until it is full.
• Refit the tank by turning it clockwise.
Stopping the washing cycle.
If the system detects a malfunction, a fault message appears on the display
screen (see paragraph 10.).
NOTE:
Once the problem has been solved, the washing cycle
automatically restarts.
Warnings for use.
WARNING: FOOT CRUSHING HAZARD
Pay attention to the patient and the staff during dental chair descent.
Summary of Contents for SMART TOUCH R7
Page 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Page 2: ......
Page 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Page 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Page 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Page 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Page 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Page 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Page 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Page 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Page 90: ......