62
62
EN
R7
- OPERATING INSTRUCTION
6.6.
Hydraulic saliva ejecto
r
The hydraulic saliva ejector starts running when the tube is removed from
the support.
Cleaning after each use.
Aspirate about ½ litre of STER 3 PLUS (CEFLA s.c.) diluted in a 6% solution
(equivalent to 60 ml of product in 1 litre of water).
Cleaning the saliva ejector filter
This operation must be carried out at the end of each work day.
WARNING!
Put on gloves before attempting to perform this operation!
• Aspirate about ½ litre of STER 3 PLUS diluted in a 6% solution (equivalent
to 60 ml of product in 1 litre of water).
• In order to prevent possible dripping of liquids and secretions from the
filter (
b
) to be extracted, aspirate only air for about 5 seconds.
• Take off the cap (
a
) by turning and twisting at the same time.
• Remove the filter (
b
).
• Clean/replace the filter (code 97290060).
• Put the filter and cap back in place.
Routine maintenance
Lubricate the o-rings (
c
) with
S1 – Protection for o-rings lubricant
.
Summary of Contents for SMART TOUCH R7
Page 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Page 2: ......
Page 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Page 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Page 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Page 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Page 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Page 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Page 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Page 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Page 90: ......