Up/Down arrow key or use the keypad and select the appropriate soft key
for minutes, seconds, or milliseconds.
q
Time Span
— Time Span sets the total duration of measurements. This
value can be set from 1 minute to 7 days. A setting of zero (MANUAL )
can also be entered if no limit on the time span is desired.
q
Auto Scale Data
— Auto Scale Data sets the reference level and scale fac-
tor to values that spread the data vertically over as much of the screen as
possible, with the reference level set in 0.1 dB steps and vertical scaling
set in 1 dB steps from 1 dB to 15 dB per division.
q
Log Data Off/On
— Log Data Off/On selects if data will be automatically
logged. When data logging is on, information is stored after each full
screen of data has been captured. Saved data is named by using the word
"LOG" and a dash followed by the time to the nearest second in 24-hour
clock format, such as "LOG - 14:43:01. Select the
RECALL DISPLAY
(0) key to view and recall the saved data (page 3-15).
q
Back
— Returns to the previous menu level.
q
Signal ID
— Signal Identification is used to determine, if possible, the air in-
terface standard of a signal. If the instrument is unable to determine the mod-
ulation format, it will display the estimated bandwidth of the signal.
q
Set IA Freq
— Set IA Freq is used to enter the center frequency of the in-
terfering signal to be identified. Enter the frequency using the Up/Down
arrow key or use the keypad and select the appropriate soft key for GHz,
MHz, kHz, or Hz.
q
Measure
— Pressing the
Measure
soft key starts the signal identification
process. This process takes approximately one minute to complete. At the
end of the measurement process, a dialog box displays the results. The
results are also shown in the message area at the bottom of the display.
q
IA Freq to Center
— The
IA Freq to Center
soft key moves the interference
analysis frequency to the center of the screen.
q
IA Freq As Marker M1
— The
IA Freq As Marker M1
soft key sets the inter-
ference analysis frequency to be the frequency of marker M1. This is the
easiest way to enter the measurement frequency. Refer to the Marker key
section on the next page for information on setting the markers.
q
Back
— Returns to the previous menu level.
q
Setup
— Sets the parameters for IA measurements.
q
Bandwidth
— Activates a menu that allows the resolution and video
bandwidths to be either coupled automatically to the span (Auto) or manu-
ally adjustable (Manual).
q
RBW Auto
— Sets the resolution bandwidth so that it is automatically
coupled to the span.
q
RBW Manual
— Sets the resolution bandwidth manually, independent
of the span, from 30 kHz to 1 MHz in zero span mode, 100 Hz to 1
MHz for a span greater than zero.
q
VBW Auto
— Sets the video bandwidth so that it is automatically cou-
pled to the RBW.
q
VBW Manual
— Sets the video bandwidth manually, independent of
the RBW, from 3 Hz to 1 MHz.
2-54
Chapter 2 Functions and Operations
Summary of Contents for Cell Master MT8212B
Page 3: ......
Page 5: ......
Page 222: ...12 4 Chapter 12 GPS Feature Figure 12 3 Saved GPS Location Information...
Page 231: ...Chapter 13 CDMA Measurement 13 9 Figure 13 5 CDMA Display with Markers...
Page 255: ...Chapter 18 EVDO Mode 18 5...