Step 4.
Connect the receiver cable to the output of the power splitter.
Step 5.
Adjust the power using the attenuator knobs until the receiver being tested drops
the signal.
Operation with Option 5, External Power Monitor
If Option 5 (External Power Monitor) is installed, it is possible to operate Option 5 and Op-
tion 28 simultaneously. This could be practical for repeater measurements or for high loss
cable loss measurements.
Procedure
Step 1.
On the Cell Master, press the
MODE
key, use the Up/Down arrow key to select
the
CW Signal Generator
mode and press
ENTER
.
Step 2.
Press the
MEAS/DISP
key.
Step 3.
Press the
Power Monitor External
soft key.
17-2
Chapter 17 CW Signal Generator
Figure 17-2.
Operation with Option 5, External Power Monitor