
InstaCal Module
The Anritsu InstaCal module can be used in place of discrete components to calibrate the
Cell Master. The InstaCal module can be used to perform an Open, Short and Load (OSL)
or a FlexCal calibration procedure. Calibration of the Cell Master with the InstaCal takes
approximately 45 seconds (see Calibration, page 3-2). Unlike a discrete calibration compo-
nent, the InstaCal module can not be used at the top of the tower to conduct load or inser-
tion loss measurements. The module operates from 2 MHz to 4 GHz and weighs eight
ounces.
Anritsu recommends annual verification of the InstaCal module to verify performance with
precision instrument data. The verification may be performed at a local Anritsu Service
Center or at the Anritsu factory.
Annual Verification
Anritsu recommends an annual calibration and performance verification of the Cell Master
and the OSL calibration components and InstaCal module by local Anritsu service centers.
Anritsu service centers are listed in Table 1-2 on the following page.
The Cell Master itself is self-calibrating, meaning that there are no field-adjustable compo-
nents. However, the OSL calibration components are crucial to the integrity of the calibra-
tion and therefore, must be verified periodically to ensure performance conformity. This is
especially important if the OSL calibration components have been accidentally dropped or
over-torqued.
ESD Precautions
The Cell Master, like other high performance instruments, is susceptible to ESD damage.
Very often, coaxial cables and antennas build up a static charge, which, if allowed to dis-
charge by connecting to the Cell Master, may damage the Cell Master input circuitry. Cell
Master operators should be aware of the potential for ESD damage and take all necessary
precautions. Operators should exercise practices outlined within industry standards like
JEDEC-625 (EIA-625), MIL-HDBK-263, and MIL-STD-1686, which pertain to ESD and
ESDS devices, equipment, and practices.
As these apply to the Cell Master, it is recommended to dissipate any static charges that
may be present before connecting the coaxial cables or antennas to the Cell Master. This
may be as simple as temporarily attaching a short or load device to the cable or antenna
prior to attaching to the Cell Master. It is important to remember that the operator may also
carry a static charge that can cause damage. Following the practices outlined in the above
standards will insure a safe environment for both personnel and equipment.
Mode References
The term “VNA” in reference to the Cell Master denotes cable and antenna analyzer modes.
The term “SPA” in reference to the Cell Master denotes Spectrum Analyzer mode. All
other modes are referenced individually.
1-14
Chapter 1 General Information
Summary of Contents for Cell Master MT8212B
Page 3: ......
Page 5: ......
Page 222: ...12 4 Chapter 12 GPS Feature Figure 12 3 Saved GPS Location Information...
Page 231: ...Chapter 13 CDMA Measurement 13 9 Figure 13 5 CDMA Display with Markers...
Page 255: ...Chapter 18 EVDO Mode 18 5...