Power Measurement
Required Equipment
q
MT8212B Cell Master
q
RS232 Serial Cable, Anritsu P/N 800-441
q
Power Supply, Anritsu P/N 40-168
q
High Accuracy Power Sensor Anritsu P/N PSN50
q
30 dB, 50W, bi-directional, DC-8.5 GHz, N(m)-N(f), Attenuator, 3-1010-123
Procedure
Step 1.
Connect the RS232 serial cable between the PSN50 High Accuracy Power Sen-
sor and the Cell Master serial port.
Step 2.
Connect the DC power supply to the DC input of the PSN50 sensor.
Step 3.
Press the
ON/OFF
key on the Cell Master.
Step 4.
Press the
MODE
key.
Step 5.
Use the Up/Down arrow key to select
High Accuracy Power Meter
mode and
press
ENTER
.
Zero and Cal
Step 6.
Press the
FREQ/DIST
key.
Step 7.
To enter the frequency manually using the number keys, select the
Center
soft
key. To select a known signal standard, press the
Signal Standard
soft key and
use the Up/Down arrow key to select the desired standard. The Cal Factors will
be derived for the corresponding center frequency.
8-2
Chapter 8 High Accuracy Power Meter
ATTENUATOR
MT8212B
DUT
RF
OUT
PSN50 SENSOR
DC SUPPLY
HOLD
RUN
START
CAL
AUTO
SCALE
SAVE
SETUP
RECALL
SETUP
LIMIT
MARKER
SAVE
DISPLAY
RECALL
DISPLAY
MODE
FREQ/DIST
AMPLITUDE
MEAS/DISP
SYS
ENTER
CLEAR
ESCAPE
ON
OFF
/
1
2
4
5
6
7
8
9
0
3
+
-
.
MT8212B
Cell
Master
Figure 8-2.
High Accuracy Power Meter Setup