Standard InstaCal Calibration
Step 1.
Select OSL Cal by pressing the
SYS
key, followed by the
Application Options
soft key. The currently selected calibration method is indicated at the bottom of
the status window. Use the
CAL Mode
soft key to select the OSL calibration
method.
Step 2.
Select the appropriate frequency range, as described on page 3-2.
Step 3.
Press the
START CAL
key. The message “Connect OPEN or InstaCal to RF
Out port” will appear in a message box, with the calibration type in the message
box title bar.
Step 4.
Connect the InstaCal module to the RF Out port (Figure 3-3).
Step 5.
Press the
ENTER
key. The Cell Master senses the InstaCal module and auto-
matically calibrates the unit using the OSL procedure. The calibration should
take about 45 seconds. An audible tone will sound when the calibration is com-
plete.
Step 6.
Verify that the calibration has been properly performed by checking that the
Cal
On!
message is displayed in the upper left corner of the status window.
FlexCal InstaCal Calibration
Step 1.
Select FlexCal by pressing the
SYS
key, followed by the
Application Options
soft key. The currently selected calibration method is indicated at the bottom of
the status window. Use the
CAL Mode
soft key to select the FlexCal calibration
method.
Step 2.
Press the
START CAL
key. The message “Connect OPEN or InstaCal to RF
Out port” will appear in a message box, with the calibration type in the message
box title bar.
Step 3.
Connect the InstaCal module to the RF Out port (Figure 3-3). The Cell Master
automatically sets the frequency from 25 MHz to 4000 MHz.
Step 4.
Press the
ENTER
key. The Cell Master senses the InstaCal module and auto-
matically calibrates the unit using the OSL procedure. The calibration should
take about 45 seconds. An audible tone will sound when the calibration is com-
plete.
Step 5.
Verify that the calibration has been properly performed by checking that the
FlexCal On!
message is displayed in the upper left corner of the status window.
Calibration with the Test Port Extension Cable
If a Test Port Extension Cable is to be used, the Cell Master must be calibrated with the
Test Port Extension Cable in place. Follow the same calibration procedures as above with
the OSL components or the InstaCal module in place at the end of the test port extension
cable.
Chapter 3 Getting Started
3-7