3-8 Noise Floor Verification
Chapter 3 — Performance Verification
3-18
PN: 10410-00730 Rev. D
MN469xC Series Multiport Test Set MM
3-8
Noise Floor Verification
This test verifies the noise floor performance of the VectorStar Multiport VNA system at the test port of the
MN469xC.
The Noise Floor test procedures are automated by using the VectorStar MS4640A/B Instrument Test Software,
CDROM part number 2300-531-R.
The software guides you through the calibration process and the measurement of the noise floor. The software
then computes the test results which are expressed in RMS values.
Refer to the
VectorStar MS4640A/B Series VNA 2300-531-R System Verification Software Quick Start Guide –
10410-00291
, for information about required equipment and details about running the software.
After the automated test is complete, print the test result and attached it the Test Records in
.