3-5 Demodulator
W-CDMA/HSPA+ Signal Analyzer
3-18
PN: 10580-00234 Rev. T
3GPP MG
6.
Press the
IQ Persistence
submenu key and use the keypad or rotary knob to set the
IQ Persistence to
2
. The instrument will display the constellation diagram after the
first sample, and then update the constellation diagram after the second sample.
IQ Persistence can be set as high as 48. When the maximum is reached, the first
sample is replaced.
Activating Markers
1.
Press the
Marker
main menu key to display the Marker menu.
2.
Press the
Marker
submenu key to select the appropriate marker (1 through 6).
The underlined marker number is the currently selected marker.
3.
Press the
On/Off
submenu key to activate the selected marker.
4.
Press the
Marker Table
submenu key to display the Marker table. The marker table is
displayed on the screen below the measurements table.
Codogram Setup
Codogram displays the code power levels over time. Two graphs are displayed on the screen,
the top one displays all the selected OVSF codes and the bottom one displays the selected
OVSF zoom codes.
Codogram Procedure
1.
Set the measurement frequency using one of the methods listed in the User Guide.
2.
Press the
Setup
main menu key.
3.
Press the
Scrambling Code
submenu key to select
Auto
so that the scrambling code is
automatically detected.
4.
Press the
S-CCPCH Spread
submenu key to manually set the S-CCPCH spreading. The
default S-CCPCH spreading factor of 256 will be displayed in all the views. Set the
S-CCPCH spreading factor to show accurate results.
5.
Press the
S-CCPCH Code
submenu key to enter the correct S-CCPCH code. The default
S-CCPCH code of 3 will be displayed in all the views. Set the S-CCPCH code to show
accurate results.
6.
Press the
PICH Code
submenu key to enter the correct PICH code. The default PICH
code of 16 will be displayed in all the views. Set the PICH code to show accurate results.
7.
Press the
Threshold
submenu key to manually set the Threshold level which determines
which codes are active. The default value is –30dB.
8.
Press the
Measurements
main menu key.
9.
Press the
Demodulator
submenu key to list the Demod menu.
Note
The W-CDMA modulation type is QPSK, and the HSPA+ modulation is 16QAM or
QPSK.
Note
Markers can be used to read the individual code power, symbol EVM (@ EVM)
and type of code and can be activated in all the W-CDMA/HSPA+ measurements.
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com
Summary of Contents for Spectrum Master MS2720T
Page 2: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 5: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 22: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 241: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 264: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 265: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 558: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 800: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1027: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1033: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1034: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1035: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...