Transmission Measurement Procedure
Step 1.
Connect the filter between the Site Master
RF In
and
RF Out
ports.
Step 2.
Set the start and stop frequencies to cover the frequency range of the filter.
Step 3.
Disconnect the duplexer and connect the cables together. If possible use all
adapters and cables that are needed to connect the duplexer to the Site Master.
Step 4.
Calibrate the Site Master for transmission measurements as described in the pre-
vious section.
Step 5.
Reconnect the duplexer and make the measurements.