3-3 Cable & Antenna Analyzer
Quick Start Guide
3-6
PN: 10580-00252 Rev. K
Site Master User Guide
Calibrate with OSL Calibration
1.
Press the
Freq/Dist
main menu key and enter the appropriate frequency range
2.
Press
Shift
, followed by
Calibrate
(2).
3.
Select
Standard
or
FlexCal
.
4.
Press
Start Cal
and follow instructions on screen.
5.
Connect Open to
RF Out
and press the
Enter
key.
6.
Connect Short to
RF Out
and press the
Enter
key.
7.
Connect Load to
RF Out
and press the
Enter
key.
8.
Verify that the calibration has been properly performed by checking that the Cal Status
message is now displaying “
ON, Standard
” or “
ON, FlexCal
”.
Note
Refer to the Cable & Antenna Measurement Guide listed in
for
calibration details.
Figure 3-5.
Calibration Setup with OSL Cal
Power
Charge
+/-
.
0
3
Sweep
2
Calibrate
1
Preset
6
Limit
5
Trace
4
Measure
9
Mode
8
System
7
File
Shift
Esc
Enter
Menu
Site
Master
S362E
OPEN
LOAD
SHORT
CALIBRATION
RFOUT/REFLECTION
TEST PORT
TEST PORT CABLE (OPTIONAL)
Summary of Contents for S331E
Page 2: ......
Page 36: ...2 9 Tilt Bail Stand Instrument Overview 2 22 PN 10580 00252 Rev K Site Master User Guide ...
Page 48: ...3 4 Spectrum Analyzer Quick Start Guide 3 12 PN 10580 00252 Rev K Site Master User Guide ...
Page 82: ...7 1 Overview Bias Tee Option 10 7 2 PN 10580 00252 Rev K Site Master User Guide ...
Page 100: ...A 1 Introduction Measurement Guides A 2 PN 10580 00252 Rev K Site Master User Guide ...
Page 124: ...Index 4 PN 10580 00252 Rev K Site Master User Guide ...
Page 125: ......