4-8
MT9090A/MU909011A Network Master User’s Guide
Figure 4-5: Parameters Setup screen
3.
Use the
Up
or
Down Arrow key
to highlight the
End of Fiber (dB)
field, if not already highlighted,
and then use the
Left
and
Right Arrow keys
to
enter the desired value. Values equal to or above this
setting are reported in the Event table.
The range for this parameter is 1.0 dB to 5.0 dB in
0.1 dB increments.
4.
Highlight the
Event Loss (dB)
field and use the
Left
and
Right Arrow keys
to enter the desired
value. Values equal to or above this setting are
reported in the Event table.
The range for this parameter is 0.25 dB to 2 dB in
0.25 dB increments.
5.
Highlight the
Reflectance (dB)
field and use the
Left
and
Right Arrow keys
to enter the desired
value. Values equal to or above this setting are
reported in the Event table.