![Anritsu MG3702xA Series Maintenance Manual Download Page 71](http://html.mh-extra.com/html/anritsu/mg3702xa-series/mg3702xa-series_maintenance-manual_2945427071.webp)
Chapter 3 — Performance Verification
3-9 Power Level Accuracy and Flatness Tests
MG3702xA MM
PN: 10370-10372 Rev. B
3-29
Power Level Flatness
For the following test, refer to the test record table in
,
“Section 3-9 Power Level Flatness”
Power level flatness is tested by measuring the output power level variation during a full band sweep in the
manual sweep mode.
Test Setup
1.
Calibrate the power meter with the appropriate power sensor.
2.
Connect the power sensor to the
RF Output
of the MG3702xA.
3.
Connect the special AUX I/O interface cable (Anritsu Part No. 806-97) to the MG3702xA rear panel
AUX I/O
connector. Connect the cable BNC connectors as follows:
a.
Connect the cable labeled “SEQ SYNC” to the power meter rear panel
INPUT 1 DIGITAL
connector.
b.
Connect the cable labeled “HORIZ OUT” to the power meter rear panel
INPUT 2 ANALOG
connector.
c.
Connect a BNC cable from the BNC tee on the INPUT 2 ANALOG connector to the DMM via
a BNC to Banana adapter.
4.
Set up the power meter as follows:
a.
Reset the power meter by pressing:
System
| Setup | -more- | PRESET | RESET
b.
Configure the power meter to perform power measurements by pressing:
Sensor
| Setup | MODE
until
Default
is displayed.
Note
Before performing this procedure, ensure that all test equipment is calibrated. Refer to the
manufacturer's test equipment manual.
Figure 3-12.
Equipment Setup for Flatness Tests and Maximum Leveled Power Test
DMM
SEQ SYNC
P/N 806-97
AUX
I/O
Power
Sensor
INPUT 2
ANALOG
INPUT 1
DIGITAL
HORIZ OUT
MG3702xA
MG37022A
Signal
Generator
Standby
Operate
ML2437A/38A
Attenuator
(If power > +19 dBm)
Summary of Contents for MG3702xA Series
Page 4: ......
Page 5: ......
Page 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Page 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Page 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Page 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Page 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Page 183: ......