Chapter 3 — Performance Verification
3-9 Power Level Accuracy and Flatness Tests
MG3702xA MM
PN: 10370-10372 Rev. B
3-23
3.
Press
Frequency
to open the current frequency parameter for editing.
4.
Press
Edit Frequency
and set F0 to the CW frequency indicated in the test record.
5.
Press
Edit Level
to open the current power level parameter for editing.
6.
Set L0 to the first applicable power level indicated in the test record.
7.
Measure the output power level with the power meter and record the reading in
,
“Section 3-9 Power Level Accuracy and Flatness Tests”
.
8.
On the MG3702xA, use the cursor control key (Arrow keys) to decrement L0 to the next test power
level in the test record. Measure and record the power meter reading in the test record.
9.
for each of the test power levels listed in the test record for the current CW
frequency.
10.
Repeat
for all CW frequencies listed in the test record.
11.
For models with external pulse modulation:
a.
Press
Modulation
to open the Pulse menu.
b.
Select external pulse mode by pressing
External
, if required.
c.
Turn the pulse mode on by pressing
On/Off
.
d.
Repeat
.
Figure 3-9.
Ensuring 0dB attenuation setting
Attenuation set
to 0 dB
Summary of Contents for MG3702xA Series
Page 4: ......
Page 5: ......
Page 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Page 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Page 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Page 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Page 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Page 183: ......