Recalling
Setups
To recall a previously saved setup, first access the
Setups menu as described below:
Step 1.
Press
System
to display the System
menu.
Step 2.
Press
Setups
to display the Setups
menu.
Step 3.
At the Setups menu, press
Recall
[RSN(M
1-9
)]
, then enter the setup number
on the keypad.
The MG369XB resets itself to the recalled configura-
tion.
Erasing
Stored Setups
The front panel setups are stored in non-volatile
memory. A master reset is required to erase the con-
tents of the setups and reprogram them with default
data.
To perform a master reset, proceed as follows:
Step 1.
With the MG369XB in standby mode,
press and hold the
RF OUTPUT ON/OFF
key.
Step 2.
Press the
LINE OPERATE/STANDBY
key
to turn the instrument on.
Step 3.
When the first menu is displayed (after
the start-up display), release the
RF OUT-
PUT ON/OFF
key.
The contents of non-volatile memory have now been
erased and reprogrammed with factory default data.
A master reset can also be performed by sending the
GPIB command “MR” over the GPIB bus.
3-86
MG369XB OM
Saving/Recalling Instrument Setups
Local (Front Panel) Operation
NOTE
The master reset function over-
writes all information stored in the
non-volatile memory with default
values. This includes the table of
2000 frequency/power level sets
used for the list sweep mode and
the five power-offset tables used for
the user level flatness correction
function.
Summary of Contents for MG369 Series
Page 3: ......
Page 4: ......
Page 14: ...1 2 MG369XB OM Figure 1 1 Series MG369XB Synthesized Signal Generator ...
Page 20: ......
Page 22: ......
Page 38: ......
Page 154: ......
Page 156: ......
Page 158: ......
Page 174: ......
Page 214: ......
Page 216: ......
Page 230: ......
Page 232: ......
Page 260: ......
Page 262: ......
Page 268: ......