3-4 Commands
Command Directory
3-88
PN: 10370-10374 Rev. F
MG369xC GPIB PM
PH8
Description:
Turns on the internal
M function in wide mode.
Disables the external
M function or the internal
M function in Narrow mode, had
either of these modes been previously programmed. If the
M function is not installed,
this command produces a syntax error.
Functional
Category:
MODULATION
Related
Commands:
PH0
,
PH7
Front Panel Key:
Modulation
| PM
(internal)
| More > | Wide
then
Modulation
| PM | On/Off
PHD <value> <unit>
Description:
Opens the internal
M deviation parameter.
Functional
Category:
PARAMETER ENTRY
Value:
0.0025 to 5 radians in Phase Modulation Narrow mode;
0.25 to 500 radians in Phase Modulation Wide mode
Units:
RD
Example:
PHD 0.5 RD; PH7
Turns internal phase modulation narrow on with a phase deviation of 500 mrad.
Output Command:
OPHD
(in rads)
Front Panel Key:
Modulation
| PM
(internal)
| Edit Deviation
PHN
Description:
Turns on the external
M function in narrow mode (Same as PH1).
Disables the internal
M function or the external
M function in Wide mode, had either
of these modes been previously programmed. If the
M function is not installed, this
command produces a syntax error.
Functional
Category:
MODULATION
Related
Commands:
PH0
,
PH2
Front Panel Key:
Modulation
| PM
(external)
| More > | Narrow
then
Modulation
| PM | On/Off