Command Directory
3-4 Commands
MG369xC GPIB PM
PN: 10370-10374 Rev. F
3-39
FDV <value> <unit>
Description:
Opens the internal FM deviation parameter.
Changes the internal FM deviation value.
Functional
Category:
PARAMETER ENTRY
Value:
10 kHz to 20 MHz in Locked, Locked Low-Noise, and Unlocked Narrow FM;
100 kHz to 100 MHz in Unlocked Wide FM
Units:
MH, KH
Example:
FM9; FDV 500 KH
Turns on FM locked mode with a deviation of 500 kHz.
Output Command:
OFD
(in Hz)
Related
Commands:
FM7
,
FM8
,
FM9
,
FMN
Front Panel Key:
Modulation
| FM (internal) | Edit Deviation
FLW <value> <data>
Description:
Prepares the signal generator to receive an FM waveform.
A 2-byte value indicating the size of the waveform (minus 1) and 2-byte values containing
the waveform data must immediately follow the command. Each waveform value is a
16-bit, unsigned integer consisting of two bytes, the upper half of the value followed by
the lower half of the value. The size of the waveform must be a power of 2.
Serial command: 0x04ab
Functional
Category:
MODULATION
Example:
FLW <2-byte binary data for waveform size><2-byte binary data for
waveform data>
Related
Commands:
FMO, FR0, FR1, WFS
Front Panel Key:
NA
FM0
Description:
Turns off the internal or external FM function.
Functional
Category:
MODULATION
Front Panel Key:
Modulation
|
FM
(internal or external) |
On/Off