Appendix B — Specifications
B-2
Performance Specifications
Pulse Power Sensor
Table B-5.
Pulse Power Sensor
Parameter/
Model
Specification
Compatible with ML2480 (with Option 15) and ML2490 Series
Power Meters
Frequency Range
MA2411A/B
(
a
)
a. MA241x sensors must use the ML2480 power meter with Option 15, 1 GHz
Calibrator, installed.
300 MHz to 40 GHz
Dynamic Range
CW: –20 dBm to +20 dBm
SWR
<1.15; 300 MHz to 2.5 GHz
<1.35; 2.5 GHz to 26 GHz
<1.5; 26 GHz to 40 GHz
Rise Time
(b)
b. Rise Time is defined as the time interval necessary for the power sensor (when
used with ML2490A series power meter) to rise from 10% to 90% of the reading
when the signal rises instantaneously from zero (no power) to 10 dBm at 25 ºC.
8 ns typical, 12 ns maximum
Sensor Linearity
<4.5% 300 MHz to 18 GHz
<7% 18 GHz to 40 GHz
RF Connector
(c)
c. Each power sensor incorporates a precision RF connector with a hexagon
coupling nut for use with an industry standard torque wrench.
Type
Pin Depth (inches)
K (m)
+0.000/–0.002
Maximum Input
Power
23 dBm, CW
30 dBm, 1
s peak, ±20 Vdc
Temperature
Accuracy
(d)
d. 10 ºC to 45 ºC
<2.0%