Appendix A Specifications
A-14
A.3.3 Time Setting
Table A.3.3-1 Time Setting
Item
Specifications
Recalculate Option
Specifies the reference parameter
Symbol Rate (Symbol Rate is automatically set when Clock Rate is
entered)
Clock Rate (Clock Rate is automatically set when Symbol Rate is entered)
Symbol Rate
Specifies the baud rate of the signal input to the Scope channel
100 Mbaud to 60 Gbaud, 1 kbaud step
Clock Rate
Specifies the clock frequency input to Trigger In
100 MHz to 15 GHz, 1 kHz step
Tracking
Function that sets Symbol Rate and Divide Ratio to values set by other
options, when BERT or Clock Recovery option is installed
Acquire Clock Rate
Uptates Clock Rate by measuring the clock rate input to Trigger In
(Data Rate is automatically set according to the Divide Ratio)
Divide Ratio
Ratio of Clock and Input Data (1 to 99. However, Pulse / Coherent Eye Mode
works only when setting 1, 2, 4, 8, 16, 32, 40, 48, 64)*
1
Divide Ratio Detect
Detects Divide Ratio automatically
Precision Trigger*
2
Enable
ON/OFF (Trigger Clock ON is possible at input frequency of 2.4 GHz or
higher)
Reset
Ability to redetect the internal correction value for Precision Trigger
Alarm
Ability to display status to the user when the internal correction value
becomes inappropriate
Scale/Offset
Unit
Specifies the horizontal axis unit of the screen
UI, Time(ps)
UI On Screen
Specifies the number of UI on the screen
Eye, Coherent Eye Mode: 1 to 100 UI (1 UI step)
Pulse Mode:
1 to 65535 UI (1 UI step)
Offset
Sets time offset for trigger position
0.00 to 32768.00 UI (0.01 UI step)
Available only when Sampling Mode is Pulse.
Software Delay
Function of correcting Inter-channel phase delay
Channel A:
Display UI converted to ps (0.1 ps step)
Channel B:
Display UI converted to ps (0.1 ps step)
Channel C:
Display UI converted to ps (0.1 ps step)
Channel D:
Display UI converted to ps (0.1 ps step)
*1: 48 is supported by Version 6 or later.
*2: Available when MP2110A-024 is installed
±
UI On Screen
2
±
UI On Screen
2
±
UI On Screen
2
±
UI On Screen
2
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...