6.5 Setting
Rate
6-67
6
How to Operate Sampling Scope
6.5.1
Setting Symbol Rate
This section explains how to set the symbol rate of the signal to measure
with Scope.
1. Click
Time.
2. Click
Rate tab.
3. Click the Tracking button of the Data Clock Rate to set to Off.
4. Click the Recalculate Option button to select Clock Rate.
5. Click the Divide Rate text box and input the division ratio.
6. Click the Symbol Rate text box and input the symbol rate.
The clock rate is calculated by the division rate and symbol rate.
Set the symbol rate and division rate so that the clock rate will be 15
000 000 kHz or less.
When BERT (MP2110A-011, MP2110A-012, or MP2110A-014) is installed,
the bit rate value of PPG can be set to the symbol rate of the sampling
oscilloscope using the following procedure.
1. Click
Time.
2. Click
the
Rate tab.
3. Click the Tracking button of the Data Clock Rate to select one of the
following.
Symbol Rate: PPG, Divide Ratio: Clock Output
Symbol Rate: PPG, Divide Ratio: Sync Output
Symbol Rate: PPG, Divide Ratio: User Defined
When Symbol Rate: PPG, Divide Ratio: User Defined is selected, set
Divide Ratio.
4. Connect
the
Trigger Clk In
connector of Scope and a BERT connector.
When Symbol Rate: PPG, Divide Ratio: User Defined is selected,
input clock with the divided frequency of
Clk Out
,
Data Out
, or
Data
Out
signal from BERT to the
Trigger Clk In
connector of Scope.
Trigger Clk In
2Vp-p Max
Scope
Status Fail
Clk Out
BERT
Output 1 2 3 4
Ext Clk In
1.6Vp-p Max
Sync Out
Sync Out
Data In
Data Out
Data Out
Data In
PPG
ED
1Vp-p Max
Ch 1
Error 1 2 3 4
Status
Fail
Bit Rate: PPG, Divide Ratio: Clock Output Is Selected
Trigger Clk In
2Vp-p Max
Scope
Status Fail
Clk Out
BERT
Output 1 2 3 4
Ext Clk In
1.6Vp-p Max
Sync Out
Sync Out
Data In
Data Out
Data Out
Data In
PPG
ED
1Vp-p Max
Ch 1
Error 1 2 3 4
Status Fail
Bit Rate: PPG, Divide Ratio: Sync Output Is Selected
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...