Chapter 6 How to Operate Sampling Scope
6-66
Setting Rate
6.5
To collect data, the trigger clock synchronized with the input signal is
required.
When BERT is installed in MP2110A, the synchronizing clock (Sync Out)
can be used as the trigger clock.
When MP2110A-054 Clock Recovery or MP2110A-055 26G/53Gbaud
Clock Recovery (SM Optical) is installed, the clock recovery unit can be
set.
After the trigger clock is input to the MP2110A, set the symbol rate, clock
rate, and divide ratio.
The symbol rate is the modulation rate of the signal to measure with
Scope (
Ch A In
or
Ch B In
, or for quad channel, the signal input to
A
,
B
,
C
or
D
).
The clock rate is the frequency of the trigger clock.
Divide Ratio is the ratio between data symbol rate and clock rate.
The value should be inputted to complete the following formula.
Symbol Rate = Clock Rate
Divide Ratio
CAUTION
● The impedance of the Trigger Clk In connector is 50 Ω.
Measurement may not be performed correctly if a cable
with another impedance is used.
● The amplitude of the signal input to the Trigger Clk In
connector is 2 Vp-p max. This is equivalent to + 10 dBm
for a sine-wave signal. Inputting a signal with a larger
voltage risks damaging the internal circuit
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...