Chapter 9 Performance Test
9-22
9.3.2 Amplitude accuracy
(1) Specification
(–2%–offset amplitude accuracy) to (+2%+offset amplitude accuracy)
The offset amplitude accuracy after calibration is as shown in the
following figure.
Figure 9.3.2-1 Offset Amplitude Accuracy
(2) Setup
Coaxial Cable
+
–
Pulse Pattern Generator
DC Power Supply
Figure 9.3.2-2 Amplitude Accuracy Test Connection Diagram