Chapter 9 Performance Test
9-16
9.2.4 Maximum input level and patterns
(1) Specification
800 mVp-p
BER 10
–
12
or less under the following conditions:
Pattern:
PRBS31, PRBS23, PRBS20, PRBS15, PRBS9,
PRBS7
Mark
ratio: 50%
Single-end
Back-to-back
connection
Without the Option 092: 11.32 Gbit/s
With the Option 092:
12.5 Gbit/s
(2) Setup
PPG1 Data Out
ED1 Data In Coaxial Terminator
Coaxial Terminator
Figure 9.2.4-1 ED1 Data In Pattern Test Connection Diagram
PPG1 Data Out
ED1 Data In
Coaxial Terminator
Coaxial Terminator
Figure 9.2.4-2 ED1
Data
In Pattern Test Connection Diagram