Chapter 9 Performance Test
9-2
9.1 Performance Test for Pulse Pattern Generator
This section describes the performance test items for the pulse pattern
generator.
Operation frequency range
Frequency accuracy
CAUTION
Connect a 50
Ω
terminator to unused input/output
connectors.
9.1.1 Required equipment
The following table shows the equipment required for the performance
test.
Before starting performance tests, warm up the BERTWave and the
measuring instruments for at least 1 hour.
Table 9.1.1-1 Equipment for Performance Test
Product Name
Required Performance
Recommended
Equipment
Sampling
oscilloscope
Electrical interface
Bandwidth: 20 GHz or more
MP2100B
(Anritsu)
Frequency counter
Frequency range:
500 MHz to 20 GHz
Accuracy:
0.1 ppm max.
MF2412C
(Anritsu)
9.1.2 Frequency accuracy
(1) Specification
±10 ppm (One hour after power on)