54100A Series
Network Analyzers
Application Note
Using Frequency Multipliers to 110 GHz
Page 1: ...54100A Series Network Analyzers Application Note Using Frequency Multipliers to 110 GHz...
Page 2: ...loss Figure 2 identifies return loss error magnitude due to source harmonics In transmission measurements sub harmonics and crossing spurs from excessive input source harmonics can make filter tuning...
Page 3: ...01pts LEVEL 7 0 dBm Average 16 2 READOUT 1 60 68 dBr 2 26 40 dBr 10 7554 GHz 1 Figure 3 This return loss plot of a tunable load illustrates high directivity and sensitivity of the measurement No inter...
Page 4: ...est accuracy when measuring low insertion loss devices 1 Select the 54100A output power to 7 0 dBm 2 Press the EXTERNAL LEVELING front panel key then 3 Press SELECT to activate the automatic level cal...