A-50
37XXXC OM
C
ALPHABETICAL LISTING
MENU
DESCRIPTION
GPIB COMMAND
CHANGE
CAL METHOD
AND LINE TYPE
None
NEXT CAL STEP
Select next calibration step. Must move cursor to here after
making below selections. Pressing the Enter key then
moves you to the next step.
None
CAL METHOD
None
STANDARD
(NOT USED FOR
WAVEGUIDE)
This option and the ones below allow you to select the
method (procedure) to be used to calibrate. This method is
independent of the calibration type, which may be 12 term,
reflection only etc.
SCM; CMX?
OFFSET SHORT
Selects offset-short method.
OCM; CMX?
LRL/LRM
Selects LRL or LRM method.
LCM; CMX?
TRM
Selects TRM method.
TCM; CMX?
TRANSMISSION
LINE TYPE
None
COAXIAL
Selects coaxial cable as the transmission line type.
LTC; LTX?
WAVEGUIDE
Selects waveguide as the transmission line type.
LTW; LTX?
MICROSTRIP
Selects microstrip as the transmission line type.
LTU; LTX?
PRESS <ENTER>
TO SELECT
Pressing the Enter key implements your menu selection.
None
Menu C11A, Select Calibration Method