Perform test as described below.
Step 1.
Reset the VNA using the Default Program key.
Step 2.
Connect Test Port 1 and Test Port 2 (top left) to-
gether.
Step 3.
Press
Ch 1
key.
Step 4.
Press
Trace Memory
key.
Step 5.
Choose VIEW DATA from menu and press
Enter
key.
Step 6.
While observing sweep indicators, allow at least two
complete sweeps to occur.
Step 7.
Choose STORE DATA TO MEMORY from menu
and press
Enter
key.
Step 8.
Choose VIEW DATA
/
MEMORY from menu and
press
Enter
key.
Step 9.
While observing sweep indicators, allow at least two
complete sweeps to occur.
Step 10.
Verify that the peak-to-peak High Level Noise falls
within the area between the two limit lines (Figure
12-2).
Step 11.
Press
Ch 3
key.
Step 12.
Repeat steps 4 thru 9 for channel 3.
Step 13.
Press
S Params
key; set
Ch 1
for S
11
and
Ch 3
for
S
22
.
Step 14.
Connect a Short to Test Port 1 and an Open to Test
Port 2 (left).
Step 15.
Repeat steps 2 thru 9.
37XXXC OM
12-7
OPERATIONAL CHECKOUT
HIGH LEVEL NOISE TEST
P OR T 1
P OR T 2
TE S T P OR T 1
TEST P OR T 2
P O R T 1
P O R T 2
S H OR T
O P E N
TEST
PORT 1
TEST
PORT 2
Summary of Contents for 37 C Series
Page 3: ......
Page 4: ......
Page 18: ...Figure 1 1 Model 37XXXC Vector Network Analyzer System...
Page 32: ......
Page 41: ......
Page 61: ...Intentionally Blank...
Page 97: ......
Page 103: ......
Page 223: ......
Page 242: ...37XXXC OM 9 21 9 22 TIME DOMAIN MENU FLOWCHART Figure 9 25 Time Domain Menu Flow...
Page 276: ......
Page 323: ......
Page 567: ......