INSTRUCTIONS FOR CONNECTING

ANRITSU 3671V50-X CABLES TO ANRITSU 37000 SERIES VNA

These cables (Figure 1) extend the VNA Port 1 and
Port 2 connectors such that the test port is effectively
m o v e d t o t h e e n d s o f t h e 3 6 7 1 V 5 0 - X c a b l e .

Instructions for connecting these cable are provided
below.

Required Equipment

1/2 inch wrench (Anritsu 01-202 or equivalent
thin 1/2-inch open-end wrench)

3/4-inch open-end wrench

Step 1.

Using the 1/2-inch wrench, carefully re-
move the test ports from Port 1 and Port
2 (Figure 2). The ports are labeled
“34YV50.”

Step 2.

Very carefully install the 34YV50 ports to
the male end of the 3671V50 cable (no vis-
ible center conductor). Hold the cable us-
ing a 3/4-inch wrench, and tighten the the
34YV50 using the 1/2-inch wrench (Figure
3). Do not over-torque.

P/N: 10410-00239

REVISION A

PRINTED: FEBRUARY 2001

COPYRIGHT 2001 ANRITSU COMPANY

490 JARVIS DRIVE

l

MORGAN HILL, CA 95037-2809

Figure 1. 3671V50-X Cables

Figure 2. Removing Test Port

3/4” wrench

1/2” wrench

Figure 3. Installing Test Port on End of Cable