3-11 Matched Thru Tests
Using the 2300-584-R Software with the ME7838E
3-28
PN: 10410-00286 Rev. L
3656B W1 Cal/Ver Kit & 2300-584-R PVS UG
3-11 Matched Thru Tests
If the
Run All Tests
button on the
Application Interface - Main Menu
tab was selected, the
Matched Thru
tests are
executed followed by the
Mismatched Thru
tests described in
Section 3-12 “Mismatched Thru Tests” on
.
If the
Matched Thru
test button was selected, only the steps in this section are executed. When this test is
complete, the program returns to the main menu described
Section 3-9 “Application Interface – Main Menu
Procedure
1.
The
Enter Serial Number Match Thru
dialog box appears.
2.
Ensure that the correct device is connected.
•
The
Matched Thru
verification standard is the device with one scribe line.
•
The
Mismatched Thru
verification standard is the device with two scribe lines.
3.
Enter the serial number of the Matched Thru.
4.
When ready to proceed, click OK. The Connect Matched Thru Male Connector dialog box appears.
5.
Connect the Matched Thru male connector to VNA Port 1 with the label facing up.
6.
Connect the Matched Thru female connector to VNA Port 2.
Note
The verification devices must be connected to the W1 Connector Coupler Ports and not to the VNA
Ports. Use the torque wrench supplied with the verification kit when tightening the device
connections.
Figure 3-39.
Anritsu W1 Verification Dialog
Note
The verification standards must be connected to the W1 Coupler Ports and not to the VNA ports.
Figure 3-40.
Matched Thru Dialog