Test Records
S312D MM
PN: 10580-00195 Revision B
A-9
Power Meter Verification (Option 29)
Calculations for power loss:
C = A – B
E = A – D
Calculations for Desired Sensor B Reading:
Desired Sensor B Reading = Test Power Level + C – E
S312D
Firmware Revision:
________________
Operator:
________________
Date:
______________
Serial Number:
________________
Options:
___________________________________________________
Table A-17.
Power Meter Verification (Option 29) Output Power Level
A
B
C
D
E
Freq
(MHz)
Sensor A
Reading at
Source Output
Sensor A
Reading at
Attenuator End
Splitter/Attenuator
Combined
Loss
Sensor B
Reading at
Power Splitter
Sensor B Path
Power Splitter
Loss
50
dBm
dBm
dBm
dBm
dBm
1000
dBm
dBm
dBm
dBm
dBm
1580
dBm
dBm
dBm
dBm
dBm
Table A-18.
Power Meter Verification (Option 29) –30 dBm Test Power Level
Frequency
Desired Sensor B Reading for –30 dBm Input
to Site Master
50 MHz
dBm
1000 MHz
dBm
1580 MHz
dBm
Table A-19.
Power Meter Verification (Option 29)
Frequency
Measured Power
Specification
50 MHz
dBm
–30.0 dBm ± 1.5 dB
100 MHz
dBm
–30.0 dBm ± 1.5 dB
1580 MHz
dBm
–30.0 dBm ± 1.5 dB
Summary of Contents for Site Master S312D
Page 2: ......
Page 10: ...1 4 Replaceable Parts General Information 1 4 PN 10580 00195 Revision B S312D MM ...
Page 48: ...3 7 RTC Battery Information Removal and Replacement 3 8 PN 10580 00195 Revision B S312D MM ...
Page 58: ...Test Records A 10 PN 10580 00195 Revision B S312D MM ...
Page 60: ...Test Fixture Schematics B 2 PN 10580 00195 Revision B S312D MM ...
Page 61: ......