Performance Verification
2-8
Spectrum Analyzer Resolution Bandwidth Accuracy Verification
S312D MM
PN: 10580-00195 Revision B
2-13
100 kHz RBW Test
1.
Press the
FREQ/DIST
key.
2.
Press the
Span
soft key, enter
150
and press the
kHz
soft key to set the span to 150 kHz.
3.
Press the
MEAS/DISP
key and the
Bandwidth
soft key.
4.
Press the
RBW Manual
soft key and use the
Up/Down
arrow key to select
100 kHz
. Press
ENTER
to set the
resolution bandwidth to 100 kHz and press
BACK
.
5.
Press the
Measure
soft key and then the
MARKER
key.
6.
Press the
M2
soft key. Record the
Δ
2 frequency in
30 kHz RBW Test
1.
Press the
FREQ/DIST
key.
2.
Press the
Span
soft key, enter
45
and press the
kHz
soft key to set the span to 45 kHz.
3.
Press the
MEAS/DISP
key and the
Bandwidth
soft key.
4.
Press the
RBW Manual
soft key and use the
Up/Down
arrow key to select
30 kHz
. Press
ENTER
to set the
resolution bandwidth to 30 kHz and press
BACK
.
5.
Press the
Measure
soft key to measure the bandwidth, then press the
MARKER
key.
6.
Press the
M2
soft key. Record the
Δ
2 frequency in
10 kHz RBW Test
1.
Press the
FREQ/DIST
key.
2.
Press the
Span
soft key, enter 15 and press the
kHz
soft key to set the span to 15 kHz.
3.
Press the
MEAS/DISP
key and the
Bandwidth
soft key.
4.
Press the
RBW Manual
soft key and use the
Up/Down
arrow key to select
10 kHz
. Press
ENTER
to set the
resolution bandwidth to 10 kHz and press
BACK
.
5.
Press the
VBW Manual
soft key and use the
Up/Down
arrow key to select
30 Hz
. Press
ENTER
to set the
video bandwidth to 30 Hz and press
BACK
.
6.
Press the
Measure
soft key, then the
MARKER
key.
7.
Press the
M2
soft key. Record the
Δ
2 frequency in
3 kHz RBW Test
1.
Press the
FREQ/DIST
key.
2.
Press the
Span
soft key, enter
4.5
and press the
kHz
soft key to set the span to 4.5 kHz.
3.
Press the
MEAS/DISP
key and the
Bandwidth
soft key.
4.
Press the
RBW Manual
soft key and use the
Up/Down
arrow key to select
3 kHz
. Press
ENTER
to set the
resolution bandwidth to 3 kHz and press
BACK
.
5.
Press the
Measure
soft key to measure the bandwidth, then press the
MARKER
key.
6.
Press the
M2
soft key. Record the
Δ
2 frequency in
1 kHz RBW Test
1.
Press the
FREQ/DIST
key.
2.
Press the
Span
soft key, enter
1.5
and press the
kHz
soft key to set the span to 1.5 kHz.
3.
Press the
MEAS/DISP
key and the
Bandwidth
soft key.
Summary of Contents for Site Master S312D
Page 2: ......
Page 10: ...1 4 Replaceable Parts General Information 1 4 PN 10580 00195 Revision B S312D MM ...
Page 48: ...3 7 RTC Battery Information Removal and Replacement 3 8 PN 10580 00195 Revision B S312D MM ...
Page 58: ...Test Records A 10 PN 10580 00195 Revision B S312D MM ...
Page 60: ...Test Fixture Schematics B 2 PN 10580 00195 Revision B S312D MM ...
Page 61: ......