6.2.3 Water supply for condenser
(only for machines equipped with this device)
This description does not concern machines with an air condenser.
If the chamber is equipped with a water condenser, adequate plumbing must be prepared with the characteristics described
in the table below.
TOWER WATER
WELL WATER
Pressure
bar
(1bar: Pa x 10
5
)
min
max
3
6
Temperature
°
C
min
max
25
30
Pressure
bar
(1bar: Pa x 10
5
)
min
max
3
6
Temperature
°
C
min
max
10
18
Connections diameter
inlet/outlet
1/2”G
Regularly check the hydraulic connections and pay special attention to the hose pipes.
The connection pipes between the water plug and the chamber must be
hose pipes in order to allow partial movement of the chamber itself during
maintenance.
Installation is advisable for regular checks of: thermometer, pressure
gauge with pressure reducer and manual on/off valves.
Pour the demineralized water directly
into storage tank 3 (fig. 6.4).
• open the lower hatch door
• unscrew the cap
• use a funnel to pour the water into
the connection pipe; the water level
can be controlled through the trans-
parent container.
Characteristics of demineralized wa-
ter:
- maximum conductivity 10
µ
S/cm.
Daily consumption:
-10 kg/24h maximum
6.4
The system is equipped with a sensor to control the water level: if this level should fall below the set limit, a warning
message will appear on the display.