background image

DN:KD-752-SMSY01 V1.0                                                                    Operation Guide 

 

11/19 

f. 

After  measurement,  the  monitor  will  turn  off  automatically  after  1  minute  with  no 

operation.  Alternatively,  you  can  press  the 

“START”  button  to  turn  off  the  monitor 

manually. 

g. 

During  measurement,  you  can  press  the 

“START”  button  to  turn  off  the  monitor 

manually. 

Note: Please 

consult a health care professional for interpretation of pressure measurements.

 

             

   

   

   

 

  Picture 7-1                  Picture 7-2                Picture 7-3                Picture 7-4 

 

3

MULTIPLE MODE

 

a. 

After applying the cuff and your body is in a comfortable position, place the standard / 

multiple switch on the 

“MULT” side. 

b. 

Press the 

“START” button, the initial information and settings are the same as 

section 7(2)b~d. 

c. 

Then multiple measurements are automatically taken and the result is automatically 

analysed and displayed. The result will also be stored in the Memory bank of the 

monitor automatically. 

d. 

After measurements, the monitor will turn off automatically after 1 minute with no 

operation. Alternatively, you can press the 

“START” button to turn off the monitor 

manually. 

e. 

The measurement result will be shown after multiple measurements. The individual 

results will not be displayed. 

f. 

Please keep quiet and calm during the entire process and do not remove the cuff 

between measurements. 

g. 

During the entire measurement process, you can press the 

“START” button to turn 

off the monitor manually. 

 

8.  DISPLAYING STORED RESULTS 

 

a. In Standby  Mode

, press “MEM” button, the monitor will display sign of current group. 

The amount of results in current user memory zone will be displayed. 

See picture 8-1

Press 

“START”  button  to  switch  group,  press  “MEM”  to  confirm  current  group.  The 

current bank can also be confirmed automatically after 5 seconds with no operation. 

b. Then LCD will display the average value of all results in the current bank. 

See picture 

8-2.

 

If no result stored in the current user memory zone, LCD will display “0” for blood 

pressure and pulse rate. 

See picture 8-3

c. 

Press “MEM” button, the LCD will display the average value of the last three results in 

the current user memory zone. 

See picture 8-4

. If no result stored in the current user 

Summary of Contents for KD-752

Page 1: ...DN KD 752 SMSY01 V1 0 Operation Guide History Revision NO Revision date Revision Description Versi on Drafter Approver 01 2018 10 25 Originated V1 0 Zhang RuoShang Zhang Fei...

Page 2: ...D 752 SMSY01 V1 0 Operation Guide Operation Guide V1 0 Product project name Wireless Blood Pressure Monitor Model name KD 752 Project number 2017 007 XY13 Drafted by Date Reviewed by Date Approved by...

Page 3: ...PROCEDURES 6 1 BATTERY LOADING 6 2 CLOCK AND DATE ADJUSTMENT 7 3 VOICE SETTING 8 4 CONNECTING THE CUFF TO THE MONITOR 9 5 APPLYING THE CUFF 9 6 BODY POSTURE DURING MEASUREMENT 10 7 TAKING YOUR BLOOD P...

Page 4: ...value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00 AM and reaches to highest level in the daytime while most people are awake...

Page 5: ...od pressure and pulse rate The most recent 4 60 measurements can be stored in the memory with date and time stamp The Electronic Sphygmomanometer corresponds to the below standards IEC 60601 1Edition...

Page 6: ...m More than 60 incl 5 12 Environmental temperature for operation 10 40 50 104 13 Environmental humidity for operation 85 RH 14 Environmental temperature for storage and transport 20 50 4 122 15 Enviro...

Page 7: ...please see part ELECTROMAGNETIC COMPATIBILITY INFORMATION It is suggested that the wireless blood pressure monitor be kept 10 meters away from other wireless devices such as WLAN unit cell phone micr...

Page 8: ...2 Consult your physician before using the device for any of the following conditions common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation arterial sclerosis poor per...

Page 9: ...OCK AND DATE ADJUSTMENT a Once you install the battery it will enter Clock Mode and LCD will display time and date See picture 2 1 The monitor will enter Standby Mode after 10 seconds with no operatio...

Page 10: ...00 12 00 PM 1 00 1 00 AM 13 00 1 00 PM 2 00 2 00 AM 14 00 2 00 PM 3 00 3 00 AM 15 00 3 00 PM 4 00 4 00 AM 16 00 4 00 PM 5 00 5 00 AM 17 00 5 00 PM 6 00 6 00 AM 18 00 6 00 PM 7 00 7 00 AM 19 00 7 00 PM...

Page 11: ...ttached to the monitor when it is packaged Should the cuff become unattached align the two plugs and four brackets of the cuff with the plug sockets and bracket sockets of the monitor and press the cu...

Page 12: ...more reliable than that from only one measurement Slide the standard multiple switch on the right side of the monitor toward to 1 to select standard mode and toward to MULT to select multiple mode be...

Page 13: ...peration Alternatively you can press the START button to turn off the monitor manually e The measurement result will be shown after multiple measurements The individual results will not be displayed f...

Page 14: ...no result which is measured from 18 o clock to 20 o clock in last 7 days stored in the current user memory zone the LCD will display 0 See picture 8 9 f Press MEM button again the most recent result w...

Page 15: ...o operation You can also press the START button to turn off the monitor manually 10 DELETING MEASUREMENTS FROM THE MEMORY When any result is displaying exclude average value displaying keeping on pres...

Page 16: ...d Apply the cuff correctly and try again Body posture was not correct during testing Review the BODY POSTURE DURING MEASUREMENT sections of the instructions and re test Speaking arm or body movement a...

Page 17: ...sure sensor parameter error No response when you press button or load battery Incorrect operation or strong electromagnetic interference Take out batteries for five minutes and then reinstall all batt...

Page 18: ...soft cloth squeezed well after moistened with water diluted disinfectant alcohol or diluted detergent 10 No component can be maintained by user in the monitor The circuit diagrams component part lists...

Page 19: ...INFORMATION Only charge the cost of components and transport SERVICE CENTER ANDON HEALTH CO LTD No 3 Jinping Street YaAn Road Nankai District Tianjin 300190 China Tel 86 22 60526081 iHealthLabs Europ...

Page 20: ...magnetic environment RF emissions CISPR 11 Group 1 Class B Home healthcare environment Harmonic distortion IEC 61000 3 2 Class A Home healthcare environment Voltage fluctuations and flicker IEC 61000...

Page 21: ...ide 19 19 810 800 960 Pulse modulation 18Hz 28V m 870 930 1720 1700 1990 Pulse modulation 217Hz 28V m 1845 1970 2450 2400 2570 Pulse modulation 217Hz 28V m 5240 5100 5800 Pulse modulation 217Hz 9V m 5...

Reviews: