DN:KD-752-SMSY01 V1.0                                                                    Operation Guide 

 

10/19 

 

Stay quiet, calm for 5 minutes before blood pressure measurement. 

 

Please keep the cuff clean. Clean the cuff by wet soft cloth and mild detergent 

if  the  cuff  becomes  dirty.  Do  not  remove  the  cuff  from  the  monitor.Clean  the 

cuff after the usage of every 200 times is recommended. 

 

Do not place the cuff around your wrist if the wrist has any inflammation, acute 

diseases, infections skin wounds. 

       

 

6.  BODY POSTURE DURING MEASUREMENT 

 

Sitting Comfortably Measurement 

a. 

Be seated with your feet flat on the floor, and don

’t cross your legs. 

b. 

Place palm upside in front of you on a flat surface such as a desk or 

table. 

c. 

The middle of the cuff should be at the level of the right atrium of 

the heart. 

 

7.  TAKING YOUR BLOOD PRESSURE READING 

 

1

SELECT STANDARD OR MULTIPLE MODE 

This  device  enables  you  to  select  either  standard  or  multiple 

mode. The standard mode is standard single measurement, and 

the  multiple  mode  is  automatic  triple  measurement.  Since  blood  pressure  constantly 

fluctuates, the blood pressure reading from multiple measurements is much more reliable 

than that from only one measurement. 

Slide the standard / multiple switch on the right side of the monitor toward to 

”1” to select 

standard mode, and toward to 

“MULT” to select multiple mode before measuring blood 

pressure.   

 

2

STANDARD MODE 

a. 

After applying the cuff and your body is in a comfortable position, place the standard / 

multiple switch on the 

“1” side. 

b. 

Press  the 

“START”  button.  All  display  characters  are  shown  for  self-test.  You  can 

check the LCD display according to the picture. Please contact the service center if 

segment is missing. 

See picture 7-1

c. 

Then  the  current  memory  bank  is  blink. 

See  picture  7-2.

  Press 

“MEM”  button  to 

change over to other bank. Confirm your selection by pressing 

“START” button. The 

current bank can also be confirmed automatically after 5 seconds with no operation. 

d. 

Then the monitor starts to seek zero pressure. 

See picture 7-3

e. 

Then the cuff will be slowly inflated. The blood pressure and pulse will be measured 

during inflation. Inflation will stop as soon as the blood pressure and pulse rate have 

been calculated and  displayed on  the  LCD. The  irregular heartbeat symbol (if any) 

and blood pressure classification indicator will blink on the LCD. 

See picture 7-4

. The 

result will automatically be stored in the Memory bank of the monitor. 

Summary of Contents for KD-752

Page 1: ...DN KD 752 SMSY01 V1 0 Operation Guide History Revision NO Revision date Revision Description Versi on Drafter Approver 01 2018 10 25 Originated V1 0 Zhang RuoShang Zhang Fei...

Page 2: ...D 752 SMSY01 V1 0 Operation Guide Operation Guide V1 0 Product project name Wireless Blood Pressure Monitor Model name KD 752 Project number 2017 007 XY13 Drafted by Date Reviewed by Date Approved by...

Page 3: ...PROCEDURES 6 1 BATTERY LOADING 6 2 CLOCK AND DATE ADJUSTMENT 7 3 VOICE SETTING 8 4 CONNECTING THE CUFF TO THE MONITOR 9 5 APPLYING THE CUFF 9 6 BODY POSTURE DURING MEASUREMENT 10 7 TAKING YOUR BLOOD P...

Page 4: ...value usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00 AM and reaches to highest level in the daytime while most people are awake...

Page 5: ...od pressure and pulse rate The most recent 4 60 measurements can be stored in the memory with date and time stamp The Electronic Sphygmomanometer corresponds to the below standards IEC 60601 1Edition...

Page 6: ...m More than 60 incl 5 12 Environmental temperature for operation 10 40 50 104 13 Environmental humidity for operation 85 RH 14 Environmental temperature for storage and transport 20 50 4 122 15 Enviro...

Page 7: ...please see part ELECTROMAGNETIC COMPATIBILITY INFORMATION It is suggested that the wireless blood pressure monitor be kept 10 meters away from other wireless devices such as WLAN unit cell phone micr...

Page 8: ...2 Consult your physician before using the device for any of the following conditions common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation arterial sclerosis poor per...

Page 9: ...OCK AND DATE ADJUSTMENT a Once you install the battery it will enter Clock Mode and LCD will display time and date See picture 2 1 The monitor will enter Standby Mode after 10 seconds with no operatio...

Page 10: ...00 12 00 PM 1 00 1 00 AM 13 00 1 00 PM 2 00 2 00 AM 14 00 2 00 PM 3 00 3 00 AM 15 00 3 00 PM 4 00 4 00 AM 16 00 4 00 PM 5 00 5 00 AM 17 00 5 00 PM 6 00 6 00 AM 18 00 6 00 PM 7 00 7 00 AM 19 00 7 00 PM...

Page 11: ...ttached to the monitor when it is packaged Should the cuff become unattached align the two plugs and four brackets of the cuff with the plug sockets and bracket sockets of the monitor and press the cu...

Page 12: ...more reliable than that from only one measurement Slide the standard multiple switch on the right side of the monitor toward to 1 to select standard mode and toward to MULT to select multiple mode be...

Page 13: ...peration Alternatively you can press the START button to turn off the monitor manually e The measurement result will be shown after multiple measurements The individual results will not be displayed f...

Page 14: ...no result which is measured from 18 o clock to 20 o clock in last 7 days stored in the current user memory zone the LCD will display 0 See picture 8 9 f Press MEM button again the most recent result w...

Page 15: ...o operation You can also press the START button to turn off the monitor manually 10 DELETING MEASUREMENTS FROM THE MEMORY When any result is displaying exclude average value displaying keeping on pres...

Page 16: ...d Apply the cuff correctly and try again Body posture was not correct during testing Review the BODY POSTURE DURING MEASUREMENT sections of the instructions and re test Speaking arm or body movement a...

Page 17: ...sure sensor parameter error No response when you press button or load battery Incorrect operation or strong electromagnetic interference Take out batteries for five minutes and then reinstall all batt...

Page 18: ...soft cloth squeezed well after moistened with water diluted disinfectant alcohol or diluted detergent 10 No component can be maintained by user in the monitor The circuit diagrams component part lists...

Page 19: ...INFORMATION Only charge the cost of components and transport SERVICE CENTER ANDON HEALTH CO LTD No 3 Jinping Street YaAn Road Nankai District Tianjin 300190 China Tel 86 22 60526081 iHealthLabs Europ...

Page 20: ...magnetic environment RF emissions CISPR 11 Group 1 Class B Home healthcare environment Harmonic distortion IEC 61000 3 2 Class A Home healthcare environment Voltage fluctuations and flicker IEC 61000...

Page 21: ...ide 19 19 810 800 960 Pulse modulation 18Hz 28V m 870 930 1720 1700 1990 Pulse modulation 217Hz 28V m 1845 1970 2450 2400 2570 Pulse modulation 217Hz 28V m 5240 5100 5800 Pulse modulation 217Hz 9V m 5...

Reviews: