2.

 

Handling  

 

 

The usual functions (displaying force, displaying maximum, zeroing and changing units) are accessible 
by simply using the keys on the keyboard. 

2.1.

 

Battery Charging  

Connect the AC adapter to the micro-USB connector on the left side of your force gauge. The green 
LED on the front panel lights up when the battery is charging. At the end of the charge, the LED goes 
out. 

  

2.2.

 

Accessories assembly  

 

 

Attach the extension cable provided in the case to the load cell rod at the bottom of the force gauge. 

Tightening should be done by hand without applying much torque. Excessive torque or the use of pliers 

can damage the load cell. We recommend that you mount the accessories with the force gauge turned 

on to ensure that you are not applying a heavy load on the sensor when mounting the accessories. 

 

 

2.3.

 

Mounting on a frame  

On the back of the force gauge are 2 M5 threaded holes that can be used to mount the force gauge on 

an ANDILOG frame. Each ANDILOG frame is delivered with a special spacer and the fixing screws for 

this use. 

 

If you wish to use another type of frame, make sure that the screws used are no longer than 3mm. 
Although the threaded holes are blind, the use of longer screws may destroy the components on the 
electronic board inside the instrument.  

 

Summary of Contents for CENTOR FIRST II

Page 1: ...s manual CENTOR FIRST II User s manual CENTOR FIRST II ANDILOG Technologies BP62001 13845 Vitrolles Cedex France COM TEN Industries 6405 49th Street North Pinellas Park FL 33781 USA Site www andilog...

Page 2: ...4 2 Getting started 6 2 1 Battery charging 6 2 2 Mounting the Accessories 6 2 3 Mounting on a frame 6 3 Using the CENTOR FIRST 7 3 1 Starting up 7 3 2 Extinguishing 8 3 3 Measurement screen 8 3 3 1 D...

Page 3: ...h a battery charge between 30 and 80 of its autonomy It is recommended to use the original power supply provided by Andilog to charge your instrument The power supply must have the following character...

Page 4: ...emperature 20 to 45 C Relative humidity 5 to 95 non condensing Maximum altitude of use 3 000 m 1 2 6 Warranty Subject to the conditions below Andilog Technologies warrants to the purchaser that it wil...

Page 5: ...entation or offers If products are damaged during shipment notify the carrier and Andilog Technologies immediately The warranty is void in case of accident misuse or abuse Calibrations overloaded sens...

Page 6: ...hould be done by hand without applying much torque Excessive torque or the use of pliers can damage the load cell We recommend that you mount the accessories with the force gauge turned on to ensure t...

Page 7: ...evice is functional the force gauge displays the measurement screen It then displays 0 N A tare is performed when the device is started If the applied force is greater than 20 of the maximum capacity...

Page 8: ...n the force gauge is turned off The force gauge will operate with the same settings when it is turned back on unit display order 3 2 Extinguishing To turn off the Centor First II press and hold the bu...

Page 9: ...on will be preceded by the symbol A force applied in compression will be preceded by the symbol The display order can be changed by pressing the 3 3 1 Do the Zero During tests it is often necessary to...

Page 10: ...display The Centor First II remembers the display order when it is turned off 3 3 4 Auto off The automatic shutdown saves the battery charge The force gauge is automatically switched off after 15 min...

Page 11: ...using the on off key My device does not display the measurement screen but an information page with Offset in red If the offset value is greater than 20 the device cannot measure Turn off your device...

Page 12: ...User s manual CENTOR FIRST II 12 Characteristic Capacity Accuracy 0 25 of capacity Autonomy 3 p m Power supply MicroUSB 5V 1A Net weight 520 g Units N daN kg lb Sensor sampling frequency 1000 Hz...

Reviews: