English 16
Cleaning
When cleaning the device and cuff, turn off the power by pressing
button
and remove the batteries.
When the main body or cuff is dirty, wipe them fully by using a gauze or
cloth dampened with water or warm water avoiding excess water.
To prevent a risk due to infection, disinfect the main body and cuff regularly.
When disinfecting them, wipe them gently by using the gauze or dampened
cloth with local antiseptic solution then wipe the moisture off the surface by
using a dry soft cloth.
Use the following medicine to disinfect the main body and cuff.
Ethanol (70%)
Isopropanol (70%)
Chlorhexidine Gluconate Solution (0.5%)
Benzalkonium Chloride Solution (0.05%)
Sodium Hypochlorite (0.05%)
11. Technical Data
Model UM-102
Measurement method
Stethoscopy with stethoscope
Measurement range
Numerical display
Pressure:
0 - 300 mmHg
Pulse:
40 - 180 beats / minute
Pressure
bar
display Pressure: 20 - 280 mmHg
Measurement accuracy
Numerical display
Pressure: ±3 mmHg
Pulse:
±5
%
Power supply
2 x 1.5 V alkaline batteries (LR6 or AA)
Upper arm circumference
22 - 32 cm using the adult cuff
Number of measurements
Approx. 2000 measurements, when AA
alkaline batteries are used, with pressure value
of 180 mmHg at room temperature of 23°C
Classification
Internally powered ME equipment
Continuous operation mode
EMC
IEC 60601-1-2: 2007
Operating conditions
+10°C to +40°C / 15%RH to 85 %RH
800 hPa to 1060 hPa
Transport / Storage conditions -20°C to +60°C / 10%RH to 95 %RH
700 hPa to 1060 hPa